Improved outcomes with an implantable left ventricular assist system: a multicenter study.

P M Portner1, P G Jansen, P E Oyer, D R Wheeldon, N Ramasamy.   

Abstract

BACKGROUND: Cumulative experience with the Novacor wearable electric left ventricular assist system (LVAS) now exceeds 850 recipients. The pump inflow conduit (IFC) has been implicated in embolic complications.
METHODS: Clinical outcomes were compared for two IFC designs in a retrospective, nonrandomized, multicenter study. The original IFC (woven, unsupported, crimped polyester: control group) and an alternative IFC (knitted, gelatin-sealed, integrally supported, uncrimped polyester: test group) were utilized upon availability in North American and European centers. Differences in cerebral embolism to 180 days postimplant were analyzed.
RESULTS: Four hundred ninety patients implanted between August 1996 and August 1999, were studied. Two hundred eighty-eight received the control IFC and 202 received the test IFC. The groups (control, test) were well matched for age (48, 49 years), etiology (idiopathic 53%[152 of 288], 55% [112 of 202]; ischemic 34% [97 of 288], 33% [66 of 202]) and mean observation time (97, 91 days). The incidence of embolic cerebrovascular accidents (CVA) was 21% (60 of 288) in the control and 12% (24 of 202) in the test group (p = 0.010). Independent risk factors for embolic CVA were found to be preimplant acute myocardial infarction (odds ratio 4.3), age above 50 years (odds ratio 2.1), and ischemic etiology (odds ratio 1.7). There was no difference in survival between the groups (71% [205 of 288], 68% [137 of 202]).
CONCLUSIONS: The alternative (test) IFC has significantly reduced the incidence of embolic CVA. This improvement is likely due to increased resistance to deformation at implant, improved neointimal adhesion, and more favorable blood flow characteristics within the conduit.

Entities:  

Mesh:

Year:  2001        PMID: 11216747     DOI: 10.1016/s0003-4975(00)02514-5

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  11 in total

Review 1.  Mechanical circulatory support devices (MCSD) in Japan: current status and future directions.

Authors:  Setsuo Takatani; Hikaru Matsuda; Akihisa Hanatani; Chisato Nojiri; Kenji Yamazaki; Tadashi Motomura; Katsuhiro Ohuchi; Tohru Sakamoto; Takashi Yamane
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

Review 2.  Development of mechanical circulatory support devices at the University of Tokyo.

Authors:  Yusuke Abe; Takashi Isoyama; Itsuro Saito; Shuichi Mochizuki; Minoru Ono; Hidemoto Nakagawa; Noriyuki Taniguchi; Norihiko Mitsumune; Ayaka Sugino; Mie Mitsui; Koki Takiura; Toshiya Ono; Akimasa Kouno; Tsuneo Chinzei; Shinichi Takamoto; Kou Imachi
Journal:  J Artif Organs       Date:  2007-06-20       Impact factor: 1.731

3.  Automatic calibration of the inlet pressure sensor for the implantable continuous-flow ventricular assist device.

Authors:  Wei Shi; Itsuro Saito; Takashi Isoyama; Hidemoto Nakagawa; Yusuke Inoue; Toshiya Ono; Akimasa Kouno; Kou Imachi; Yusuke Abe
Journal:  J Artif Organs       Date:  2011-03-05       Impact factor: 1.731

4.  The helical flow pump with a hydrodynamic levitation impeller.

Authors:  Yusuke Abe; Kohei Ishii; Takashi Isoyama; Itsuro Saito; Yusuke Inoue; Toshiya Ono; Hidemoto Nakagawa; Emiko Nakano; Kyoko Fukazawa; Kazuhiko Ishihara; Kazuyoshi Fukunaga; Minoru Ono; Kou Imachi
Journal:  J Artif Organs       Date:  2012-08-28       Impact factor: 1.731

5.  Japanese multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system.

Authors:  Ryozo Omoto; Shunei Kyo; Motonobu Nishimura; Hikaru Matsuda; Goro Matsumiya; Soichiro Kitamura; Takeshi Nakatani; Shinichi Takamoto; Mdminoru Ono; Koichi Tabayashi; Ryohei Yozu
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

6.  A reliable test to detect impending pump failure during long-term support on the Novacor N100 Left Ventricular Assist System.

Authors:  P M Dohmen; H Laube; K de Jonge; W Konertz
Journal:  Tex Heart Inst J       Date:  2001

7.  Incidence and patterns of adverse event onset during the first 60 days after ventricular assist device implantation.

Authors:  Elizabeth A Genovese; Mary Amanda Dew; Jeffrey J Teuteberg; Marc A Simon; Joy Kay; Michael P Siegenthaler; Jay K Bhama; Christian A Bermudez; Kathleen L Lockard; Steve Winowich; Robert L Kormos
Journal:  Ann Thorac Surg       Date:  2009-10       Impact factor: 4.330

8.  Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices.

Authors:  Danny Bluestein; Gaurav Girdhar; Shmuel Einav; Marvin J Slepian
Journal:  J Biomech       Date:  2012-12-06       Impact factor: 2.712

Review 9.  Mechanical circulatory support devices for acute heart failure syndromes: considerations for clinical trial design.

Authors:  John B O'Connell; Patrick M McCarthy; George Sopko; Gerasimos S Filippatos; Ileana L Piña; Marvin A Konstam; James B Young; Leslie W Miller; Mandeep R Mehra; Edmond Roland; John E A Blair; David J Farrar; Mihai Gheorghiade
Journal:  Heart Fail Rev       Date:  2008-06-12       Impact factor: 4.214

10.  HeartWare miniature axial-flow ventricular assist device: design and initial feasibility test.

Authors:  Mark S Slaughter; Michael A Sobieski; Daniel Tamez; Timothy Horrell; Joel Graham; Patroklos S Pappas; Antone J Tatooles; Jeffrey LaRose
Journal:  Tex Heart Inst J       Date:  2009
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.