| Literature DB >> 11209456 |
R S Marks1, A Margalit, A Bychenko, E Bassis, N Porat, R Dagan.
Abstract
A chemiluminescent-based optical fiber immunosensor was developed for the detection of antipneumococcal antibodies. This was accomplished by developing a different chemical procedure utilizing 3-aminopropyl trimethoxysilane and cyanuric chloride to conjugate pneumococcal cell wall polysaccharides to the optical fiber tips, and by improving the sensitivity of the photodetection system. The lowest titer of antipneumococcal antibodies detected by the optical fiber was at a 1:819,200 dilution. The lowest corresponding value by standard enzyme-linked immunosorbent assay was at a 1:98,415 dilution. It was concluded that the optical immunosensor system is an accurate and sensitive method to detect antipneumococcal antibodies and may be an adequate tool to monitor antibodies in specimens such as saliva and urine.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11209456 DOI: 10.1385/abab:89:2-3:117
Source DB: PubMed Journal: Appl Biochem Biotechnol ISSN: 0273-2289 Impact factor: 2.926