Optimization of background electrolytes for capillary electrophoresis I. Mathematical and computational model.

B Gas1, P Coufal, M Jaros, J Muzikár, I Jelínek.   

Abstract

A mathematical and computational model is introduced for optimization of background electrolyte systems for capillary zone electrophoresis of anions. The model takes into account mono- or di- or trivalent ions and allows also for modeling of highly acidic or alkaline electrolytes, where a presence of hydrogen and hydroxide ions is significant. At maximum, the electrolyte can contain two co-anions and two counter-cations. The mathematical relations of the model are formulated to enable an easy algorithmization and programming in a computer language. The model assesses the composition of the background electrolyte in the analyte zone, which enables prediction of the parameters of the system that are experimentally available, like the transfer ratio, which is a measure of the sensitivity in the indirect photometric detection or the molar conductivity detection response, which expresses the sensitivity of the conductivity detection. Furthermore, the model also enables the evaluation of a tendency of the analyte to undergo electromigration dispersion and allows the optimization of the composition of the background electrolyte to reach a good sensitivity of detection while still having the dispersion properties in the acceptable range. Although the model presented is aimed towards the separation of anions, it can be straightforwardly rearranged to serve for simulation of electromigration of cationic analytes. The suitability of the model is checked by inspecting the behavior of a phosphate buffer for analysis of anions. It is shown that parameters of the phosphate buffer when used at neutral and alkaline pH values possess singularities that indicate a possible occurrence of system peaks. Moreover, if the mobility of any analyte of the sample is close to the mobilities of the system peaks, the indirect detector signals following the background electrolyte properties will be heavily amplified and distorted. When a specific detector sensitive on presence of the analyte were used, the signal would be almost lost due to the excessive dispersion of the peak.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11206794     DOI: 10.1016/s0021-9673(00)00983-3

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  3 in total

1.  Electrophoretic Behavior of Anionic Triazine and PAMAM Dendrimers: Methods for Improving Resolution and Assessing Purity Using Capillary Electrophoresis.

Authors:  Sanjiv Lalwani; Vincent J Venditto; Abdellatif Chouai; Gregory E Rivera; Sunil Shaunak; Eric E Simanek
Journal:  Macromolecules       Date:  2009-04-28       Impact factor: 5.985

2.  Mimicking PAMAM Dendrimers with Ampholytic, Hybrid Triazine Dendrimers: A Comparison of Dispersity and Stability.

Authors:  Sanjiv Lalwani; Abdellatif Chouai; Lisa M Perez; Vanessa Santiago; Sunil Shaunak; Eric E Simanek
Journal:  Macromolecules       Date:  2009-08-12       Impact factor: 5.985

3.  High-Throughput Analysis of Lidocaine in Pharmaceutical Formulation by Capillary Zone Electrophoresis Using Multiple Injections in a Single Run.

Authors:  Andressa C Valese; Daniel A Spudeit; Maressa D Dolzan; Lizandra C Bretanha; Luciano Vitali; Gustavo A Micke
Journal:  J Anal Methods Chem       Date:  2016-03-16       Impact factor: 2.193

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.