Depth of cure and compressive strength of dental composites cured with blue light emitting diodes (LEDs).

K D Jandt1, R W Mills, G B Blackwell, S H Ashworth.   

Abstract

OBJECTIVE: The primary objective of this pilot study was to test the hypotheses that (i) depth of cure and (ii) compressive strength of dental composites cured with either a light emitting diode (LED) based light curing unit (LCU) or a conventional halogen LCU do not differ significantly. The second objective of this study was to characterise irradiance and the emitted light spectra for both LCUs to allow comparisons between the units.
METHODS: Dental composite (Spectrum TPH, shades A2 and A4) was cured for 40 s with either a commercial halogen LCU or a LED LCU, respectively. The LED LCU uses 27 blue LEDs as the light source. The composites' depth of cure was measured for 10 samples of 4 mm diameter and 8 mm depth for each shade with a penetrometer. The results were compared using a Student's t-test. Compressive strengths were determined after 6 and 72 h, for six samples of 4 mm diameter and 6 mm depth for each shade after being polymerised for 40 s from each end of the mould. Groups were compared using a three way ANOVA.
RESULTS: The conventional halogen LCU cured composites significantly (p < 0.05) deeper (6.40 mm A2, 5.19 mm A4) than did the LED LCU (5.33 mm A2, 4.27 mm A4). Both units cured the composite deeper than required by both ISO 4049 and the manufacturer. A three way ANOVA showed that there were no significant differences in the compressive strengths of samples produced with either the LED LCU or the halogen LCU (p = 0.460). Significant differences in compressive strength of samples stored for 6 and 72 h (p = 0.0006) and of samples of different shades (p = 0.035) were found as confirmed by the three way ANOVA. The light spectra of both units differed strongly. While the halogen LCU showed a broad distribution of wavelengths with a power peak at 497 nm, the LED LCU emitted most of the generated light at 465 nm. The LED LCU produced a total irradiance of 350 mW cm-2 whereas the halogen LCU produced a total irradiance of 755 mW cm-2. SIGNIFICANCE: The results showed that both units provided sufficient output to exceed minimum requirements in terms of composites' depth of cure according to ISO 4049 and the depth of cure and the composites' compressive strength stated by the manufacturer. Compressive strengths of dental composites cured under laboratory conditions with a LED LCU were statistically equivalent to those cured with a conventional halogen LCU. With its inherent advantages, such as a constant power output over the lifetime of the diodes, LED LCUs have great potential to achieve a clinically consistent quality of composite cure.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11203522     DOI: 10.1016/s0109-5641(99)00083-4

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  33 in total

1.  Influence of the LED curing source and selective enamel etching on dentin bond strength of self-etch adhesives in class I composite restorations.

Authors:  Eduardo José Souza-Junior; Cíntia Tereza Pimenta Araújo; Lúcia Trazzi Prieto; Luís Alexandre Maffei Sartini Paulillo
Journal:  Lasers Med Sci       Date:  2011-11-26       Impact factor: 3.161

2.  The effect of curing light and chemical catalyst on the degree of conversion of two dual cured resin luting cements.

Authors:  Eduardo José Souza-Junior; Lúcia Trazzi Prieto; Giulliana Panfiglio Soares; Carlos Tadeu dos Santos Dias; Flávio Henrique Baggio Aguiar; Luís Alexandre Maffei Sartini Paulillo
Journal:  Lasers Med Sci       Date:  2010-11-23       Impact factor: 3.161

3.  Effect of power density of curing unit, exposure duration, and light guide distance on composite depth of cure.

Authors:  Anders Lindberg; Anne Peutzfeldt; Jan W V van Dijken
Journal:  Clin Oral Investig       Date:  2005-04-07       Impact factor: 3.573

4.  Evaluation of the curing depth of two translucent composite materials using a halogen and two LED curing units.

Authors:  Olga Polydorou; Alexandros Manolakis; Elmar Hellwig; Petra Hahn
Journal:  Clin Oral Investig       Date:  2007-10-25       Impact factor: 3.573

5.  Temperature rise and degree of photopolymerization conversion of nanocomposites and conventional dental composites.

Authors:  Mohammad Atai; Fariba Motevasselian
Journal:  Clin Oral Investig       Date:  2008-12-16       Impact factor: 3.573

6.  Characterization of heat emission of light-curing units.

Authors:  Mohammed A Wahbi; F A Aalam; F I Fatiny; S A Radwan; I Y Eshan; K H Al-Samadani
Journal:  Saudi Dent J       Date:  2012-03-09

7.  Efficiency of light-emitting diode and halogen units in reducing residual monomers.

Authors:  Felipe de Assis Ribeiro Carvalho; Rhita C Almeida; Marco Antonio Almeida; Lucia H S Cevidanes; Marcia C Amorim M Leite
Journal:  Am J Orthod Dentofacial Orthop       Date:  2010-11       Impact factor: 2.650

8.  Clinical long-term success of contemporary nano-filled resin composites in class I and II restorations cured by LED or halogen light.

Authors:  Torsten Pflaum; Stefan Kranz; Regina Montag; Arndt Güntsch; Andrea Völpel; Robin Mills; Klaus Jandt; Bernd Sigusch
Journal:  Clin Oral Investig       Date:  2017-10-28       Impact factor: 3.573

9.  Evaluation of depth of cure and knoop hardness in a dental composite, photo-activated using different methods.

Authors:  Mithra N Hegde; Priyadarshini Hegde; Babita Malhan
Journal:  J Conserv Dent       Date:  2008-04

10.  Cytotoxicity and oxidative stress caused by dental adhesive systems cured with halogen and LED lights.

Authors:  Gianrico Spagnuolo; Marco Annunziata; Sandro Rengo
Journal:  Clin Oral Investig       Date:  2003-12-16       Impact factor: 3.573

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.