J D Regan1, M Sherriff, N Meredith, K Gulabivala.
Abstract
The pattern of dentine removal during endodontic instrumentation is influenced by many factors including the interfacial forces applied by the operator. The aim of this study was to investigate the influence of operators and different sizes and types of instruments on the magnitude of these interfacial forces. Single-rooted teeth were mounted on a cantilevered aluminium beam to which two pairs of single element strain gauges were joined in a half-bridge configuration and mounted at right angles to each other. The strain gauges were connected to an analogue-to-digital converter fitted in a micro-computer via conditioning amplifiers. This enabled strains to be recorded over a period of time. Twenty operators instrumented root canals using a series of hand instruments for 1 min each. The mean interfacial forces used by operators demonstrated a wide variation ranging from 9.06 g to 149.42 g (range of forces from 0-331 g) but there was a consistency in the relative magnitude for each operator. The 20 operators could be divided into 13 groups which were significantly different (alpha = 0.05) from each other. There were significant differences (alpha = 0.05) between the forces used for each of the K-Flex files (15, 25, 35, 45, and 70), the force increasing with the file size. There was also a significant difference (alpha = 0.05) in the forces used between the Flexofile (#25) and the #25 K-Flex and Hedström files. However, there was no significant difference between the K-Flex and Hedström files.Entities:
Mesh:
Year: 2000 PMID: 11202864 DOI: 10.1034/j.1600-9657.2000.016003101.x
Source DB: PubMed Journal: Endod Dent Traumatol ISSN: 0109-2502