K H Hong1, Y H Cho, M H Lee, G S Choi, C W Lee.
Abstract
Two radiostrontium isotopes, 89Sr and 90Sr, were measured simultaneously by liquid scintillation counter coupling with a full spectrum DPM counting protocol. Strontium was separated from the aqueous media by extraction chromatography using Sr-spec resin. Individual count rates were evaluated from quench correlation curves for 89Sr and 90Sr related with SIS and tSIE quench indicators. The method exhibited a counting efficiency of 95% for 89Sr and 92% for 90Sr within their full energy channels and gave highly precise results when the sample was counted within 4 h after strontium separation. Lower limits of detection were 37 mBq/l for 90Sr and 32 mBq/l for 89Sr with a 60 min counting time.Entities: