Stereotactic localization of an intracranial lesion by computed tomography or magnetic resonance imaging requires the use of a head frame that is fixed to the skull of the patient. To such head frames are attached either N-shaped or V-shaped localization rods. Because of patient positioning, the transverse imaging slices may not be parallel to the frame base; a coordinate transformation algorithm that takes this possibility into consideration is crucial. Here we propose such an algorithm for a head frame with V-shaped localization rods. Our algorithm determines the transformation matrix between the image coordinate system of a transverse image and the frame coordinate system. The determining procedure has three steps: (a) calculation of the oblique angles of a transverse image relative to the head frame and calculation of the image magnification factor; (b) determination of the coordinates of four central markers in both coordinate systems; and (c) determination of the 3 x 3 transformation matrix by using the coordinates of the four markers. This algorithm is robust in principle and is useful for improving the accuracy of localization.Entities:
Mesh:
Year: 2001 PMID: 11197682 DOI: 10.1088/0031-9155/46/1/401
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609