Accuracy of gamma probes in localizing radioactivity: in-vitro assessment and clinical implications.

P I Haigh1, E C Glass, R Essner.   

Abstract

BACKGROUND: Radioguided surgery (RGS) uses gamma probes to localize radioactive targets. We examined the in-vitro localizing properties of commercially available gamma probes for three common RGS radioisotopes: technetium-99m (99mTc), indium-111 (111In), and iodine-125 (125I).
METHODS: The linear and angular localizing properties of five gamma probes were assessed. Radioactive counts in air were obtained at multiple energy threshold settings for each radioisotope. The full width half maximum (FWHM) and angular half maximum (AHM) were calculated to determine localizing accuracy.
RESULTS: The FWHM of the five probes ranged from 1.2 to 3.4 cm for 99mTc, 1.8 to 2.7 cm for 111In and 1.6 to 2.8 cm for 125I. AHM of the probes ranged from 20 to 49 degrees for 99mTc, 14 to 35 degrees for 111In, and 18 to 47 degrees for 125I. The ranking of probe accuracy depended on the radioisotope studied. Absolute count rates varied with energy threshold settings for each probe and radioisotope. Probes required individualized energy threshold settings for optimal performance.
CONCLUSIONS: The five probes varied in their characteristics for localizing the three radioisotopes. Each probe has particular detection characteristics that may change with different radioisotopes. Differences in radioisotope detection may explain intraoperative variation in localization during lymphatic mapping and sentinel lymph node dissection (LM/SL). Surgeons should be aware of probe performance characteristics before intraoperative use.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11190488     DOI: 10.1089/cbr.2000.15.561

Source DB:  PubMed          Journal:  Cancer Biother Radiopharm        ISSN: 1084-9785            Impact factor:   3.099


  2 in total

1.  Evaluation of two intraoperative gamma detectors for assessment of 177Lu activity concentration in vivo.

Authors:  Viktor Sandblom; Ingun Ståhl; Roger Olofsson Bagge; Eva Forssell-Aronsson
Journal:  EJNMMI Phys       Date:  2017-01-09

2.  Intra-operative use of PET probe for localization of FDG avid lesions.

Authors:  M A Molina; W J Goodwin; F L Moffat; A N Serafini; G N Sfakianakis; E Avisar
Journal:  Cancer Imaging       Date:  2009-09-29       Impact factor: 3.909

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.