Mechanical reliability of the AMS 700CXM inflatable penile prosthesis for the treatment of male erectile dysfunction.

Y Deuk Choi1, Y Jin Choi, J Hwan Kim, H Ki Choi.   

Abstract

PURPOSE: The AMS 700CXM inflatable penile prosthesis with medium controlled expansion cylinders was introduced in 1990 to provide a device with controlled expansion in girth and fitness for Asian men. We reviewed our experience with the AMS 700CXM prosthesis with particular regard to postoperative complications and reliability.
MATERIALS AND METHODS: From January 1991 to April 1999, 273 men with erectile dysfunction underwent primary penile prosthesis implantation with the AMS 700CXM. Functional status of the prosthesis was assessed retrospectively by reviewing the medical records and a telephone interview.
RESULTS: Patient age was 24 to 78 years and mean followup was 49 months. Of the 28 complications (10.3%) 8 (3%) were nonmechanical and 20 (7.3%) were mechanical. Nonmechanical complications included infection in 2 cases, erosion in 3, surgical complications in 2 and tube kinking in 1. Mechanical complications involved a cylinder tear in 11 cases (4%), pump failure in 4 (1.5%) and a reservoir tear in 2 (0.7%). Mechanical reliability of the device was 98.2% after 2, 95.7% after 3 and 90.4% after 5 years.
CONCLUSIONS: The AMS 700CXM is suitable for Asian men and its mechanical reliability is excellent. However, more device modifications are needed to guarantee good reliability during long-term followup.

Entities:  

Mesh:

Year:  2001        PMID: 11176478     DOI: 10.1097/00005392-200103000-00021

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  3 in total

Review 1.  A comparative review of the options for treatment of erectile dysfunction: which treatment for which patient?

Authors:  Konstantinos Hatzimouratidis; Dimitrios G Hatzichristou
Journal:  Drugs       Date:  2005       Impact factor: 9.546

2.  Tubing erosion of an inflatable penile prosthesis long after implantation.

Authors:  Alvaro Morales
Journal:  Sex Med       Date:  2014-06       Impact factor: 2.491

Review 3.  Prevention, identification, and management of post-operative penile implant complications of infection, hematoma, and device malfunction.

Authors:  Timothy K O'Rourke; Alexander Erbella; Yu Zhang; Matthew S Wosnitzer
Journal:  Transl Androl Urol       Date:  2017-11