Sensitometric evaluation of a new F-speed dental radiographic film.

C Price1.   

Abstract

OBJECTIVE: To compare a new experimental Kodak (Eastman Kodak, Rochester, NY, USA) F-speed dental X-ray film with Ektaspeed Plus and Ultra-speed.
METHODS: The three types of film were exposed and processed under standardized conditions. Values of base plus fog, speed, and film contrast were derived. Resolution was compared by a line-pair plate.
RESULTS: Under these experimental conditions, the speed of the new film was just into the F-speed range, and a little over twice as fast as Ultra-speed. The Ektaspeed Plus emulsion was somewhat slower than previously recorded when it was first introduced, but around the centre of the E-speed range. Ektaspeed Plus and the new F-speed film had almost identical film contrasts, the F-speed film having slightly greater contrast in the higher density range. Ultra-speed contrast was marginally greater in the lower density range, but was overtaken by both of the other emulsions at higher densities. All three emulsions had low values of base plus fog. Both E- and F-speed films resolved 10 line-pairs per millimetre well, though both emulsions were inferior to Ultra-speed.
CONCLUSIONS: The new F-speed film, since marketed as Insight, provides a further saving in radiation exposure, with no evident deterioration in film contrast or resolution.

Mesh:

Substances:

Year:  2001        PMID: 11175270     DOI: 10.1038/sj/dmfr/4600569

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  2 in total

1.  Effects of developer exhaustion on DFL Contrast FV-58 and Kodak Insight dental films.

Authors:  Fabiano Pachêco de Carvalho; M M F da Silveira; M A G Frazão; S T de Santana; M L dos Anjos Pontual
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

Review 2.  Radiographic modalities for diagnosis of caries in a historical perspective: from film to machine-intelligence supported systems.

Authors:  Ann Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2021-03-04       Impact factor: 3.525

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.