Image-receptor performance: a comparison of Trophy RVG UI sensor and Kodak Ektaspeed Plus film.

J Ludlow1, A Mol.   

Abstract

UNLABELLED: Objective. This study compares the physical characteristics of the RVG UI sensor (RVG) with Ektaspeed Plus film. STUDY
DESIGN: Dose-response curves were generated for film and for each of 6 available RVG modes. An aluminum step-wedge was used to evaluate exposure latitude. Spatial resolution was assessed by using a line-pair test tool. Latitude and resolution were assessed by observers for both modalities. The RVG was further characterized by its modulation transfer function.
RESULTS: Exposure latitude was equal for film and RVG in the periodontal mode. Other gray scale modes demonstrated much lower latitude. The average maximum resolution was 15.3 line-pairs per millimeter (lp/mm) for RVG in high-resolution mode, 10.5 lp/mm for RVG in low-resolution mode, and 20 lp/mm for film (P <.0001). Modulation transfer function measurements supported the subjective assessments.
CONCLUSIONS: In periodontal mode, the RVG UI sensor demonstrates exposure latitude similar to that of Ektaspeed Plus film. Film images exhibit significantly higher spatial resolution than the RVG images acquired in high-resolution mode.

Mesh:

Year:  2001        PMID: 11174581     DOI: 10.1067/moe.2001.110574

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  1 in total

1.  The usage of digital radiography and cone beam computed tomography among Turkish dentists.

Authors:  Semanur Dölekoğlu; E Fişekçioğlu; M İlgüy; D İlgüy
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.