D J Herlihy1, D J Larkman, N M deSouza, A D Williams, I R Young.
Abstract
A coil tuned to 21.3 MHz was incorporated into a nasogastric tube and used as a marker of tube position during magnetic resonance (MR) imaging in a 0.5-T scanner. Catheter tracking was investigated with the coil used in both transmit/receive and in receive-only modes. Data acquired from this coil were overlaid on images obtained using the body coil of the scanner. Visualization of the full length of the catheter with local high signal at the tip was achieved with a temporal resolution of approximately 1 second. J. Magn. Reson. Imaging 2001;13:127-130. Copyright 2001 Wiley-Liss, Inc.Mesh:
Year: 2001 PMID: 11169814 DOI: 10.1002/1522-2586(200101)13:1<127::aid-jmri1019>3.0.co;2-k