Survey of conventional and spiral ct doses.

N Hidajat1, M Wolf, A Nunnemann, P Liersch, B Gebauer, U Teichgräber, R J Schröder, R Felix.   

Abstract

PURPOSE: To investigate the radiation dose for conventional computed tomography (CT) and spiral CT during different CT examinations at various hospitals and practices.
MATERIALS AND METHODS: CT dose index with an active length of 15 cm was measured in 16 different types of CT scanners by using ionization chamber dosimetry. Twenty-six holders (one who has legal responsibility under national law for a radiologic installation) operating a total of seven conventional and 20 spiral CT scanners were asked for their standard parameters for various CT examinations. Weighted CT dose index and dose-length product were determined for each examination.
RESULTS: For most examinations, the tube current time product was significantly higher for conventional CT than for spiral CT (.002 </= P </=.05). The ratio of section distance to section thickness for conventional CT was significantly lower than the pitch for spiral CT (.001 </= P </=.05). The weighted CT dose index and dose-length product for spiral CT were about half of those for conventional CT. The third quartiles for weighted CT dose index and dose-length product for spiral CT were much lower than those recommended as reference doses.
CONCLUSION: CT examinations with conventional CT scanners are often performed with unnecessarily high radiation dose. For the establishment of reference doses, the radiation dose with spiral CT scanners should be taken into account.

Mesh:

Year:  2001        PMID: 11161152     DOI: 10.1148/radiology.218.2.r01ja12395

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  Comparison of image quality between conventional and low-dose nonenhanced head CT.

Authors:  Mark E Mullins; Michael H Lev; Peter Bove; Cara E O'Reilly; Sanjay Saini; James T Rhea; James H Thrall; George J Hunter; Leena M Hamberg; R Gilberto Gonzalez
Journal:  AJNR Am J Neuroradiol       Date:  2004-04       Impact factor: 3.825

2.  Survey of computed tomography techniques and absorbed dose in Italian hospitals: a comparison between two methods to estimate the dose-length product and the effective dose and to verify fulfilment of the diagnostic reference levels.

Authors:  Daniela Origgi; Sabrina Vigorito; Gaetano Villa; Massimo Bellomi; Giampiero Tosi
Journal:  Eur Radiol       Date:  2005-02-12       Impact factor: 5.315

3.  Evaluation of automated attenuation-based tube current adaptation for coronary calcium scoring in MDCT in a cohort of 262 patients.

Authors:  Georg Mühlenbruch; Christian Hohl; Marco Das; Joachim E Wildberger; Christoph Suess; Ernst Klotz; Thomas Flohr; Ralf Koos; Christoph Thomas; Rolf W Günther; Andreas H Mahnken
Journal:  Eur Radiol       Date:  2007-02-17       Impact factor: 5.315

4.  Dose management in CT facility.

Authors:  V Tsapaki; M Rehani
Journal:  Biomed Imaging Interv J       Date:  2007-04-01

5.  Reducing Radiation Doses in Female Breast and Lung during CT Examinations of Thorax: A new Technique in two Scanners.

Authors:  P Mehnati; M Ghavami; H Heidari
Journal:  J Biomed Phys Eng       Date:  2017-09-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.