Power Doppler US: evaluation of the morphology of stenoses with a flow phantom.

M Claudon1, D Winninger, S Briançon, P Pesque.   

Abstract

PURPOSE: To determine the importance of technical settings at power Doppler ultrasonography (US) for the evaluation of stenoses.
MATERIALS AND METHODS: A computer-controlled phantom was used to induce a reproducible flow across a calibrated 71% stenosis in an 8.4-mm-diameter tube. Two transducers, 2-4 and 5-10 MHz, working at depths of 3.0 and 11.5 cm, respectively, with different beam angles (40 degrees, 60 degrees, and 90 degrees ), were used to simulate evaluation of pulsatile flow across normal and stenotic vessels in various conditions. For each condition, gain, pulse repetition frequency, and wall filter were progressively turned from low to high values. Two observers measured in a blinded fashion the apparent lumen of the stenotic and normal vessels on longitudinal and transverse images with the use of power Doppler US.
RESULTS: When the high-frequency transducer was used, gain significantly affected both stenotic and feeding vessel measurement, whereas pulse repetition frequency and filter only affected feeding vessel evaluation. When the low-frequency transducer was used, all factors, including flow velocity and beam angle, played a significant role (P <.001). In most conditions, overestimation of the lumen and underestimation of the lumen of the feeding vessel led to severe underestimation of the degree of stenosis.
CONCLUSION: Power Doppler US cannot be used to measure stenoses accurately. Underestimation of the degree of the stenosis was significantly higher with the low-frequency probe than with the high-frequency probe.

Mesh:

Year:  2001        PMID: 11152788     DOI: 10.1148/radiology.218.1.r01ja03109

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Diagnostics and characterisation of preocclusive stenoses and occlusions of the internal carotid artery with B-flow.

Authors:  E M Jung; R Kubale; G Ritter; M T Gallegos; K-P Jungius; N Rupp; D-A Clevert
Journal:  Eur Radiol       Date:  2006-05-16       Impact factor: 5.315

2.  Color Doppler, power Doppler and B-flow ultrasound in the assessment of ICA stenosis: Comparison with 64-MD-CT angiography.

Authors:  D-A Clevert; T Johnson; E M Jung; D-A Clevert; P M Flach; T I Strautz; G Ritter; M T Gallegos; R Kubale; C Becker; M Reiser
Journal:  Eur Radiol       Date:  2006-11-22       Impact factor: 7.034

3.  A comparison of Power Doppler with conventional sonographic imaging for the evaluation of renal artery stenosis.

Authors:  Agatino Manganaro; Giuseppe Ando'; Antonino Salvo; Andrea Consolo; Frank Coppolino; Domenico Giannino
Journal:  Cardiovasc Ultrasound       Date:  2004-01-14       Impact factor: 2.062

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.