| Literature DB >> 1115046 |
Abstract
An improved micromethod for plasma fibrinogen, based on the Ellis-Stransky thrombin-clot density procedure, is presented. Modifications includes: (1) addition of Polybrene eliminates heparin interference; (2) measurement of absorbance at 340 nm provides improved sensitivity at low fibrinogen levels as well as with small volumes of plasma (40 mul.); (3) higher concentration of thrombin reduces the incubation time. Plasma fibrinogen by the proposed method compares well with a reference procedure based on total clottable protein (coefficient of correlation 0.99).Entities:
Mesh:
Substances:
Year: 1975 PMID: 1115046 DOI: 10.1093/ajcp/63.3.426
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493