Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients.

K Tanigawa1, T Takeda, E Goto, K Tanaka.   

Abstract

BACKGROUND: To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients.
METHODS: Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambulance crews when they arrived at the authors' department. Immediately after intubation in the emergency department, the endotracheal tube position was tested by the SIB and end-tidal carbon dioxide (ETCO2) monitor using an infrared carbon dioxide analyzer. We observed the SIB reinflating for 10 s, and full reinflation within 4 s was defined as a positive result (tracheal intubation).
RESULTS: Five esophageal intubations occurred, and the SIB correctly identified all esophageal intubations. Of the 65 tracheal intubations, the SIB correctly identified 47 tubes placed in the trachea (72.3%). Delayed but full reinflation occurred in one tracheal intubation during the 10-s observation period. Fifteen tracheal intubations had incomplete reinflation during the observation period, and two tracheal intubations did not achieve any reinflation. Thirty-nine tracheal intubations were identified by ETCO2 (60%). When the SIB test is combined with the ETCO2 detection, 59 tracheal intubations were identified with a 90.8% sensitivity.
CONCLUSIONS: The authors found a high incidence of false-negative results of the SIB in out-of-hospital cardiac arrest patients. Because no single test for verifying endotracheal tube position is reliable, all available modalities should be tested and used in conjunction with proper clinical judgment to verify tracheal intubation in cases of out-of-hospital cardiac arrest.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11149438     DOI: 10.1097/00000542-200012000-00015

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  6 in total

Review 1.  Cardiopulmonary resuscitation and management of cardiac arrest.

Authors:  Jerry P Nolan; Jasmeet Soar; Volker Wenzel; Peter Paal
Journal:  Nat Rev Cardiol       Date:  2012-06-05       Impact factor: 32.419

2.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

Review 3.  Ultrasonography in Emergency Department; a Diagnostic Tool for Better Examination and Decision-Making.

Authors:  Ali Abdolrazaghnejad; Mohsen Banaie; Mohammad Safdari
Journal:  Adv J Emerg Med       Date:  2017-12-11

Review 4. 

Authors:  J P Nolan; C D Deakin; J Soar; B W Böttiger; G Smith; M Baubin; B Dirks; V Wenzel
Journal:  Notf Rett Med       Date:  2006-02-01       Impact factor: 0.826

5.  Predictive Value of Tracheal Rapid Ultrasound Exam Performed in the Emergency Department for Verification of Tracheal Intubation.

Authors:  Babak Masoumi; Reza Azizkhani; Gilava Hedayati Emam; Morteza Asgarzadeh; Behrouz Zargar Kharazi
Journal:  Open Access Maced J Med Sci       Date:  2017-06-17

6.  Point-of-Care Ultrasound for Intubation Confirmation of COVID-19 Patients.

Authors:  Michael Gottlieb; Stephen Alerhand; Brit Long
Journal:  West J Emerg Med       Date:  2020-08-17
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.