Literature DB >> 11139160

The integrity of latex gloves in clinical dental practice.

F A Pitten1, G Herdemann, A Kramer.   

Abstract

BACKGROUND: The aim of the study was to assess the integrity of non-sterile latex gloves (Biogel Diagnostic (A), Biogel Dental (B), Gentle Skin (C) and Manufix (D)) in clinical dental practice.
MATERIALS AND METHODS: After use, the gloves were collected and examined for perforations using the water inflation test according to the EN 455-1 Standard. Over a 12-month period 847 used gloves were collected and assessed. The mean duration of use for gloves A/B/C/D was 175/112/78/79 min respectively. In spite of these differences favoring gloves C and D, 16% of glove A, 14% of glove B, 21% of glove C and 29% of glove D were found to leak. The proportion of perforated gloves increased slightly if a hygienic handrub using an alcoholic hand disinfectant had been carried out immediately prior to gloving.
CONCLUSION: The results indicate striking differences between the quality of medical gloves. As all the gloves comply with the European regulations in terms of integrity, it is very difficult for the user to distinguish between gloves of superior or inferior quality. In a clinical setting it is therefore recommended that the appropriate glove should be determined for each situation. Hands must be thoroughly dry if they are disinfected prior to gloving, since hands still wet with alcoholic disinfectant may increase the risk of glove perforation.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11139160     DOI: 10.1007/s150100070011

Source DB:  PubMed          Journal:  Infection        ISSN: 0300-8126            Impact factor:   3.553


  12 in total

1.  Does a preceding hand wash and drying time after surgical hand disinfection influence the efficacy of a propanol-based hand rub?

Authors:  Nils-Olaf Hübner; Günter Kampf; Philipp Kamp; Thomas Kohlmann; Axel Kramer
Journal:  BMC Microbiol       Date:  2006-06-22       Impact factor: 3.605

2.  Suppression of surgeons' bacterial hand flora during surgical procedures with a new antimicrobial surgical glove.

Authors:  Ojan Assadian; Axel Kramer; Kenneth Ouriel; Miranda Suchomel; Mary-Louise McLaws; Martin Rottman; David Leaper; Afshin Assadian
Journal:  Surg Infect (Larchmt)       Date:  2013-10-12       Impact factor: 2.150

Review 3.  [Hygienic and dermatologic aspects of hand disinfection and prophylactic skin antisepsis].

Authors:  A Kramer; M Jünger; G Kampf
Journal:  Hautarzt       Date:  2005-08       Impact factor: 0.751

4.  Integrity of disposable nitrile exam gloves exposed to simulated movement.

Authors:  Robert N Phalen; Weng Kee Wong
Journal:  J Occup Environ Hyg       Date:  2011-05       Impact factor: 2.155

5.  [HCV, HBV and HIV infections: risk for surgeon and staff. Results and consequences of routine screening in emergency patients].

Authors:  K Dresing; C Pouwels; S Bonsack; M Oellerich; H Schwörer; A Uy; K M Stürmer
Journal:  Chirurg       Date:  2003-11       Impact factor: 0.955

6.  Bacterial migration through punctured surgical gloves under real surgical conditions.

Authors:  Nils-Olaf Hübner; Anna-Maria Goerdt; Natalie Stanislawski; Ojan Assadian; Claus-Dieter Heidecke; Axel Kramer; Lars Ivo Partecke
Journal:  BMC Infect Dis       Date:  2010-07-01       Impact factor: 3.090

7.  Antibacterial activity of a sterile antimicrobial polyisoprene surgical glove against transient flora following a 2-hours simulated use.

Authors:  Johannes Leitgeb; Rupert Schuster; Bit New Yee; Pui Fong Chee; Julian-Camill Harnoss; Peter Starzengruber; Michael Schäffer; Ojan Assadian
Journal:  BMC Surg       Date:  2015-07-04       Impact factor: 2.102

8.  Incidence and patterns of surgical glove perforations: experience from Addis Ababa, Ethiopia.

Authors:  Abebe Bekele; Nardos Makonnen; Lidya Tesfaye; Mulat Taye
Journal:  BMC Surg       Date:  2017-03-20       Impact factor: 2.102

9.  The durability of examination gloves used on intensive care units.

Authors:  Nils-Olaf Hübner; Anna-Maria Goerdt; Axel Mannerow; Ute Pohrt; Claus-Dieter Heidecke; Axel Kramer; Lars Ivo Partecke
Journal:  BMC Infect Dis       Date:  2013-05-20       Impact factor: 3.090

Review 10.  Indications and the requirements for single-use medical gloves.

Authors:  Axel Kramer; Ojan Assadian
Journal:  GMS Hyg Infect Control       Date:  2016-01-12
View more