Evaluation of the effectiveness of decontamination of dental syringes.

K Vickery1, A Pajkos, Y Cossart.   

Abstract

AIM: Steam autoclaving is the gold standard for decontaminating dental instruments, but worldwide disinfection is still widely employed. We have evaluated a range of procedures for their ability to inactivate duck hepatitis B virus contaminating dental syringes.
METHODS: Residual infectivity of virus suspensions following 2% glutaraldehyde treatment, ultrasonication or steam sterilisation at 121 degrees or 134 degrees was assayed by injecting day-old ducklings and examining their livers for viral DNA 2.5 weeks later. Dental syringes were contaminated with DHBV positive blood, then treated by the same methods. An anaesthetic cartridge containing water was loaded into the syringe and 400microl aliquots used to inject day-old ducklings. Used dental syringes were examined by Scanning Electron Microscopy.
RESULTS: Suspension test:- ultrasonic treatment failed to inactivate DHBV in suspension, but complete inactivation was achieved by 2% glutaraldehyde and autoclaving. Syringe test:- neither ultrasonic treatment nor glutaraldehyde inactivated DHBV. Autoclaving at 134 degrees (3 minutes) permitted transmission to 1/16 ducklings but steam sterilisation at 121 degrees (15 minutes) was effective. Electronmicroscopy demonstrated organic debris (biofilm) in the lumen of used syringes.
CONCLUSION: Short autoclaving cycles, albeit at raised temperatures, may fail to inactivate the virus because of poor steam penetration, inadequate heat transfer and the accumulation of protective biofilm.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11132693     DOI: 10.1038/sj.bdj.4800847

Source DB:  PubMed          Journal:  Br Dent J        ISSN: 0007-0610            Impact factor:   1.626


  4 in total

Review 1.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

2.  Susceptibility of Staphylococcus aureus biofilms to reactive discharge gases.

Authors:  Christian Traba; Jun F Liang
Journal:  Biofouling       Date:  2011-08       Impact factor: 3.209

Review 3.  Is hepatitis B-virucidal validation of biocides possible with the use of surrogates?

Authors:  Andreas Sauerbrei
Journal:  World J Gastroenterol       Date:  2014-01-14       Impact factor: 5.742

4.  Enzymatic detergent reuse in gastroscope processing: a potential source of microorganism transmission.

Authors:  Maria Letícia de Miranda Mati; Natália Rocha Guimarães; Paula Prazeres Magalhães; Luiz de Macêdo Farias; Adriana Cristina de Oliveira
Journal:  Rev Lat Am Enfermagem       Date:  2019-12-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.