Abstract
Only target-conformal treatment allows one to exploit the advantages of ion beams (the increased dose and high biological efficiency at the end of the particle range) to a maximum extent. Up to now, target-conformal treatments such as spot scanning or intensity-controlled raster scanning have used fast magnetic lateral deflection in one or two directions perpendicular to the beam axis and a slow range variation in the longitudinal axis by active or passive energy variation. The present paper describes a new method for conformal irradiation with a fast intensity-controlled longitudinal scan in the beam direction, called 'depth scanning'. Its advantages and disadvantages will be discussed. First experimental results from depth scanning will be presented.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11131189 DOI: 10.1088/0031-9155/45/12/309