Literature DB >> 1112850

Failed femoral stems in total hip prostheses. A report of six cases.

J O Galante, W Rostoker, J M Doyle.   

Abstract

Six femoral stems of total joint prostheses failed and were studied. These included two short-neck Müller, one standard Müller, one new design long-neck Müller, and two Charnley prostheses. In addition, reference is made to another failed Charnley prosthesis which had not required revision at the time of the study. Reoperation and replacement of the femoral component was required in all six cases. Metal fatigue appeared to be the cause of failure in all instances. Metallographic examination of the removed prosthesis revealed no underlying defects in one prosthesis. In the other five prostheses metallurgical defects were found. Varus placement of the femoral prosthesis or loosening of the cement-prosthesis bond, or both, were identified in five of the six and were thought to have led to overload. A combination of metal defects and loosening or malpositioning was thought to be responsible for the failures. It was further indicated that current designs should be regarded as marginal in relation to long service life, until more fatigue information is available on the metallic materials currently in use.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 1112850

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  8 in total

1.  Early failure of a non-cemented femoral stem after minimal-invasive total hip arthroplasty: cause analysis and classification.

Authors:  Jan P Benthien; Wolf Siepen
Journal:  Musculoskelet Surg       Date:  2012-05-06

2.  [Critical comment on the work of G. Ritter et al.: "Elastic strapping of hip joint prosthesis". (Arch. orthop. Unfall-Chir. 86, 1-14 (1976))].

Authors:  R Kölbel; A Rohlmann; G Bergmann; G Friedebold
Journal:  Arch Orthop Unfallchir       Date:  1977-08-26

3.  Failure of the stem in total hip replacement. A study of aetiology and mechanism of failure in 13 cases.

Authors:  U E Pazzaglia; F Ghisellini; D Barbieri; L Ceciliani
Journal:  Arch Orthop Trauma Surg       Date:  1988

4.  An analysis of 119 loosenings in total hip endoprostheses.

Authors:  P Bösch; H Kristen; K Zweymüller
Journal:  Arch Orthop Trauma Surg       Date:  1980

5.  Studies on biodegradation of crosslinked hydroxy terminated-poly(proplyene fumarate) and formation of scaffold for orthopedic applications.

Authors:  K T Shalumon; M Jayabalan
Journal:  J Mater Sci Mater Med       Date:  2008-06-27       Impact factor: 3.896

6.  Fracture hip prostheses due to welding defects.

Authors:  R Kousbroek; P Ducheyne; G Van der Perre; E Aernoudt; J C Mulier
Journal:  Arch Orthop Trauma Surg       Date:  1981

7.  Fracture of cobalt chrome, fully-coat beaded femoral revision long stem, a clinical outcomes study.

Authors:  Chih-Hsiang Chang; Po-Chun Lin; Cheng-Min Shih; Chun-Chieh Chen; Pang-Hsin Hsieh; Hsin-Nung Shih
Journal:  Biomed J       Date:  2018-03-30       Impact factor: 4.910

8.  A case of a C-stem fracture at the head-neck junction and a review of the literature.

Authors:  David Morley; Ian Starks; Justin Lim
Journal:  Case Rep Orthop       Date:  2012-12-09
  8 in total