Selective recording of electroneurograms from the sciatic nerve of a dog with multi-electrode spiral cuffs.

J Rozman1, B Zorko, M Bunc.   

Abstract

Electroneurograms (ENGs) from superficial regions of the sciatic nerve of a Beagle dog were recorded selectively with a chronically implanted 33-electrode spiral cuff (cuff). By delivering stimulating pulses to groups of three electrodes (GTEs) within the cuff we could define the relative positions of the particular superficial regions that selectively innervated the tibialis anterior (TA) and gastrocnemius muscles (GM). GTEs with and without contractions of the TA and GM muscles were selected and connected to a 4-channel ENG system designed to amplify ENGs by 100,000 times and to pass frequencies between 500 Hz and 10 kHz. In our study, 12 experiments were conducted on three Beagle dogs with a cuff implanted for up to 2 years. We present the results obtained in four experiments conducted on one animal. With the implanted leg mounted in a special electronic brace we applied extending forces to the ankle, rotating it by up to 37 degrees according to the neutral position, eliciting torque to stretch the TA muscle. Only the ENG from a GTE eliciting maximum contraction of the TA muscle showed activities corresponding to the trajectory of the mechanical load of the muscle. Next, we dissected the calcanean tendon (CT) of the implanted leg and applied repetitive pull forces to the CT. Only the ENG from the GTE eliciting maximum contraction of the GM muscle was activated in correspondence to the trajectory of the mechanical load applied on the CT. The results suggest that the cuff, implanted chronically on the sciatic nerve, is useful to record ENGs of the afferent fibers from TA and GM muscles selectively and that the technique could be extended for human use in the field of rehabilitation for paralysis.

Entities:  

Mesh:

Year:  2000        PMID: 11120917     DOI: 10.2170/jjphysiol.50.509

Source DB:  PubMed          Journal:  Jpn J Physiol        ISSN: 0021-521X


  7 in total

1.  Fabrication of High Contact-Density, Flat-Interface Nerve Electrodes for Recording and Stimulation Applications.

Authors:  Yazan M Dweiri; Matthew A Stone; Dustin J Tyler; Grant A McCallum; Dominique M Durand
Journal:  J Vis Exp       Date:  2016-10-04       Impact factor: 1.355

2.  Hierarchical beamformer and cross-talk reduction in electroneurography.

Authors:  Daniela Calvetti; Brian Wodlinger; Dominique M Durand; Erkki Somersalo
Journal:  J Neural Eng       Date:  2011-07-29       Impact factor: 5.379

3.  Localization and recovery of peripheral neural sources with beamforming algorithms.

Authors:  Brian Wodlinger; Dominique M Durand
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2009-10-16       Impact factor: 3.802

4.  Model-based Bayesian signal extraction algorithm for peripheral nerves.

Authors:  Thomas E Eggers; Yazan M Dweiri; Grant A McCallum; Dominique M Durand
Journal:  J Neural Eng       Date:  2017-07-04       Impact factor: 5.379

5.  Effect on signal-to-noise ratio of splitting the continuous contacts of cuff electrodes into smaller recording areas.

Authors:  Max Ortiz-Catalan; Jorge Marin-Millan; Jean Delbeke; Bo Håkansson; Rickard Brånemark
Journal:  J Neuroeng Rehabil       Date:  2013-02-21       Impact factor: 4.262

6.  Mapping of Small Nerve Trunks and Branches Using Adaptive Flexible Electrodes.

Authors:  Zhuolin Xiang; Swathi Sheshadri; Sang-Hoon Lee; Jiahui Wang; Ning Xue; Nitish V Thakor; Shih-Cheng Yen; Chengkuo Lee
Journal:  Adv Sci (Weinh)       Date:  2016-03-23       Impact factor: 16.806

7.  Influence of nerve cuff channel count and implantation site on the separability of afferent ENG.

Authors:  Carolina Silveira; Emma Brunton; Sally Spendiff; Kianoush Nazarpour
Journal:  J Neural Eng       Date:  2018-04-09       Impact factor: 5.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.