Literature DB >> 11119281

Entrapment of subvalvular mitral tissue causing intermittent failure of a St Jude mitral prosthesis.

F Agostini1, R L Click, S L Mulvagh, M D Abel, J A Dearani.   

Abstract

Frequently portions of the mitral valve and sub-valvular apparatus are left intact during mitral valve replacement to help preserve left ventricular function. We describe a patient with paroxysmal congestive heart failure caused by intermittent entrapment of the subvalvular apparatus in the prosthesis, preventing complete valve closure.

Entities:  

Mesh:

Year:  2000        PMID: 11119281     DOI: 10.1067/mje.2000.107251

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  1 in total

1.  Transvalvular mitral regurgitation following mitral valve replacement a diagnostic dilemma.

Authors:  U S Dinesh Kumar; Umesh Nareppa; Shyam Prasad Shetty; Murugesh Wali
Journal:  Ann Card Anaesth       Date:  2015 Oct-Dec