[Possible errors in leukocyte count caused by carryover in the Coulter counter S].

R Kuse, K Hausmann.   

Abstract

In 163 samples of venous blood with significant leukocytosis (20 000 to 172000 cells pro mm-3) a carryover of 400 to 22000 leukocytes up to the third count of the following specimens (200 to 9300 cells/mm-3) could be observed. Even after 41 specimens of normal leukocyte range and low leukocytosis (6700 to 17700 cells/mm-3) in cases of extreme leukopenias (less than 2000 cells/mm-3) the correct result was exceeded by 500 to 1000 cells. In four additional machines of other laboratories a carryover of different extent could be reproduced. In comparison to samples of venous blood only a very small carryover into subsequent samples of isotonic solution was found. The considerable spreading of the coefficient of variation between 1.7 and 16.2% (arithmetic mean 7.1%) prohibited a mathematical correction of error. Erroneously too high numbers of leukocytes could be absolutely avoided by at least two repeated counts.

Entities:  

Mesh:

Year:  1975        PMID: 1111659     DOI: 10.1007/bf01633965

Source DB:  PubMed          Journal:  Blut        ISSN: 0006-5242


  7 in total

1.  [Automation in clinical chemical analyses (author's transl)].

Authors:  R Haeckel
Journal:  Z Klin Chem Klin Biochem       Date:  1972-06

2.  [Treatment of the error due to overspill in the determination of sodium, potassiuma and chloride with the automatic analyser C 4].

Authors:  R Haeckel; A J Porth
Journal:  Z Klin Chem Klin Biochem       Date:  1972-02

3.  Evaluation of the Coulter Counter Model S.

Authors:  G M Brittin; G Brecher; C A Johnson
Journal:  Am J Clin Pathol       Date:  1969-12       Impact factor: 2.493

4.  An evaluation of the Coulter S counter.

Authors:  A A Sharp; B C Ballard
Journal:  J Clin Pathol       Date:  1970-05       Impact factor: 3.411

5.  An assessment of the Coulter counter model S.

Authors:  P H Pinkerton; I Spence; J C Ogilvie; W A Ronald; P Marchant; P K Ray
Journal:  J Clin Pathol       Date:  1970-02       Impact factor: 3.411

6.  An evaluation of the autoanalyzer SMA-4.

Authors:  T R Lappin; A Lamont; M G Nelson
Journal:  J Clin Pathol       Date:  1969-01       Impact factor: 3.411

7.  [Comparative leucocyte counting from vein, ear lobe and finger (author's transl)].

Authors:  R Kuse; K Hausmann
Journal:  Dtsch Med Wochenschr       Date:  1973-10-12       Impact factor: 0.628

  7 in total
  1 in total

1.  [Detection of reagent errors by internal quality control of the Coulter counter S].

Authors:  R Kuse; K Hausmann
Journal:  Blut       Date:  1976-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.