Is it safe to reuse disposable laparoscopic trocars? An in vitro testing.

A C Chan1, M Ip, A Koehler, B Crisp, J S Tam, S C Chung.   

Abstract

BACKGROUND: The reuse of disposable laparoscopic instruments carries a risk of transmitting infectious diseases such as hepatitis and HIV. We evaluated the safety of reusing disposable trocars by studying the chances of their harboring infectious viruses after resterilization in an in vitro setting.
METHODS: Disposable laparoscopic trocars were exposed to horse blood contaminated with high or low viral concentrations of herpes simplex virus type 1 (HSV1) and attenuated polio virus type 1 at room temperature for 2 h. HSV1 was chosen as the surrogate for lipid viruses that include hepatitis B and HIV virus; polio virus represented the nonlipid viruses that cause infections in immunocompromised patients and are more resistant to sterilization. The trocars were subsequently cleaned and resterilized by low-temperature steam and formaldehyde at 80 degrees C for 3 h. Viral cultures were then repeated after sterilization.
RESULTS: A cytopathic effect (CPE) was demonstrated at both concentrations for HSV1 in all trocars before but not after sterilization. For the polio virus, CPE was evident in 50% of the trocars (two of four) exposed to high viral concentration after sterilization.
CONCLUSION: Disposable trocars are difficult to resterilize and may harbor infectious viruses after their initial use. Therefore, the reuse of disposable trocars in laparoscopic surgery cannot be recommended.

Entities:  

Mesh:

Year:  2000        PMID: 11116415     DOI: 10.1007/s004640000118

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  3 in total

1.  Reprocessed single-use devices in laparoscopy: assessment of cost, environmental impact, and patient safety.

Authors:  David Renton; Peter Denk; Oliver Varban
Journal:  Surg Endosc       Date:  2018-06-15       Impact factor: 4.584

2.  Replacement of expensive, disposable instruments with old-fashioned surgical techniques for improved cost-effectiveness in laparoscopic hysterectomy.

Authors:  John E Morrison; Volker R Jacobs
Journal:  JSLS       Date:  2004 Apr-Jun       Impact factor: 2.172

3.  Day case laparoscopic cholecystectomy at Kilimanjaro Christian Medical Centre, Tanzania.

Authors:  Imogen Cullen; Fadlo Shaban; Oroog Ali; Matthew Breckons; Kondo Chilonga; Daudi Wapalila; Jamil Suleiman; Mercy Elinisa; Bronwyn Woodburn; Richard Walker; Liam Horgan
Journal:  Surg Endosc       Date:  2020-09-01       Impact factor: 4.584

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.