The purpose of this study was to evaluate the clinical and radiological results of unstable pelvic fractures treated with a new external fixation device. Between May 1992 and May 1998, 43 patients with unstable pelvic fractures were treated with a new anterior pelvic external fixator. Two died, and therefore 41 patients' results were evaluated. There were 29 men and 12 women, and their average age was 34 years (range 12-70 years). Traffic accidents accounted for 34 injuries. Three patients fell from a height, 3 were injured in industrial accidents, and 1 was hit by a train. According to the Tile classification, there were 24 type B pelvic injuries and 17 type C. Associated injuries were observed in 21 patients. A considerable reduction of the pelvic pain was noted after application of the fixator in all patients. Excessive blood transfusion was not required in any patient. The average follow-up was 24 months (range 12-50 months). Clinical results at final evaluation were good according to the criteria of Matta and Saucedo in 34 patients and poor in 7. In conclusion, the new pelvic external fixator is effective, safe, and easy to apply in the treatment of unstable pelvic fractures. The fixator can be used alone in patients with type B pelvic injuries such as open book and lateral compression. However, it does not provide sufficient stability for severely displaced type C injuries when applied alone. Nevertheless, it may be helpful for fixing type C injuries like a posterior iliac fracture without dislocation of the sacroiliac joint.

Entities:  

Mesh:

Year:  2000        PMID: 11110141     DOI: 10.1007/s004020000142

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  4 in total

1.  13-Year experience in external fixation of the pelvis: complications, reduction and removal.

Authors:  P M Mitchell; C M Corrigan; N A Patel; A J Silverberg; S E Greenberg; R V Thakore; W T Obremskey; J M Ehrenfeld; J M Evans; M K Sethi
Journal:  Eur J Trauma Emerg Surg       Date:  2015-02-24       Impact factor: 3.693

2.  Complications of anterior subcutaneous internal fixation for unstable pelvis fractures: a multicenter study.

Authors:  Rahul Vaidya; Erik N Kubiak; Patrick F Bergin; Derek G Dombroski; Ren J Critchlow; Anil Sethi; Adam J Starr
Journal:  Clin Orthop Relat Res       Date:  2012-08       Impact factor: 4.176

3.  Midterm Radiographic and Functional Outcomes of the Anterior Subcutaneous Internal Pelvic Fixator (INFIX) for Pelvic Ring Injuries.

Authors:  Rahul Vaidya; Adam Jonathan Martin; Matthew Roth; Frederick Tonnos; Bryant Oliphant; Jon Carlson
Journal:  J Orthop Trauma       Date:  2017-05       Impact factor: 2.512

Review 4.  Patient Weight-bearing after Pelvic Fracture Surgery-A Systematic Review of the Literature: What is the Modern Evidence Base?

Authors:  Mark Rickman; Bjorn-Christian Link; Lucian B Solomon
Journal:  Strategies Trauma Limb Reconstr       Date:  2019 Jan-Apr
  4 in total