G H Isaac1, C A Busch, V Shetty, K J Drabu.
Abstract
The results are reported of a radiographic study of cement mantle thickness in 112 consecutive primary hip replacements. Measurements were made by three observers of the apparent cement thickness medially and laterally using standard anterior-posterior radiographs. The average cement thickness was 3.2 mm, which is 1.2 mm greater than the size difference between the broach and the prosthesis, and was in the range 2-5 mm in 67 per cent of all measurement points. This has significance for the design of instrumentation to prepare the femoral cavity to give a defined cement mantle thickness. There was a greater cement mantle thickness proximally than distally. In 95 cases it was possible to determine the orientation of the stem within the cement mantle, which showed an even distribution between varus and valgus orientation; 49 per cent were within 1 degree of neutral and only one case was more than 5 degrees from neutral.Mesh:
Substances:
Year: 2000 PMID: 11109854 DOI: 10.1243/0954411001535499
Source DB: PubMed Journal: Proc Inst Mech Eng H ISSN: 0954-4119 Impact factor: 1.617