Influence of perfusion rate and compliance on esophageal manometry.

M P Zabinski, H M Spiro, P Biancani.   

Abstract

The accuracy of esophageal manometry using the conventional perfused three-lumen catheter depends on the proper choice of the perfusion rate. Previous investigators have used a wide range of perfusion rates without considering the compliance of the transducer-catheter assembly as a variable in the selection of the perfusion rate for the hydraulic systems. We present a relationship relating the perfusion rate and the compliance which specifies the minimum adequate perfusion rate for a given transducer-catheter assembly. This approach explains apparent discrepancies in the literature regarding the proper rate of perfusion for accurate measurement of esophageal contractions. Esophageal motility tests with 10 subjects confirmed the accuracy of manometric tests at the prescribed perfusion rate (1.25 ml/min) since they compared well with intraluminal pressures measured directly with a Honeywell intraesophageal transducer. The effect of perfusion rate on the apmlitude of esophageal contraction was also investigated and it is shown that high perfusion rates result in increased amplitudes of contraction. Furthermore, at large perfusion rates a large volume of liquid builds up in the esophagus. This fluid accumulation could possibly be responsible for the increased amplitudes of contraction.

Entities:  

Mesh:

Year:  1975        PMID: 1110235     DOI: 10.1152/jappl.1975.38.1.177

Source DB:  PubMed          Journal:  J Appl Physiol        ISSN: 0021-8987            Impact factor:   3.531


  4 in total

1.  Four-channel sleeve catheter for prolonged measurement of lower esophageal sphincter pressure.

Authors:  J H Schneider; P F Crookes; H D Becker
Journal:  Dig Dis Sci       Date:  1999-12       Impact factor: 3.199

2.  Pressure tension, and force of closure of the human lower esophageal sphincter and esophagus.

Authors:  P Biancani; M P Zabinski; J Behar
Journal:  J Clin Invest       Date:  1975-08       Impact factor: 14.808

3.  Oesophageal manometry by liquid-filled catheters.

Authors:  A Shaw; A D Baillie; J Runcie
Journal:  Med Biol Eng Comput       Date:  1980-07       Impact factor: 2.602

4.  Rhythmic contraction but arrhythmic distension of esophageal peristaltic reflex in patients with dysphagia.

Authors:  Kazumasa Muta; Ravinder K Mittal; Ali Zifan
Journal:  PLoS One       Date:  2022-01-24       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.