Literature DB >> 11089471

An isocentrically mounted stand for total body irradiation.

D Jones1, J W Rieke, B L Madsen, M D Hafermann.   

Abstract

The purpose of this study was to construct a stand to support a patient for total body photon irradiation and to expedite the set-up and treatment by rotating the stand. As in other isocentric treatments, the midline dose is impacted less by source-to-skin distance variations. The method of immobilizing the patient is described. A 10 mm lucite plate is supported in front of the patient to increase skin dose. A matrix of holes in this plate serves to index the location of blocks used to shield the lungs. The dosimetry of the set-up is described, as is the production of tissue deficit compensators. The results of phantom studies and in vivo thermoluminescent dosimetry measurements are presented.

Entities:  

Mesh:

Year:  2000        PMID: 11089471     DOI: 10.1259/bjr.73.871.11089471

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  1 in total

1.  Dosimetry and verification of Co total body irradiation with human phantom and semiconductor diodes.

Authors:  Mahmoud Allahverdi; Ghazale Geraily; Mahbod Esfehani; Aliakbar Sharafi; Peyman Haddad; Alireza Shirazi
Journal:  J Med Phys       Date:  2007-10