Literature DB >> 11066110

Transcatheter retrieval and repositioning of an Amplatzer device embolized into the left atrium.

M Peuster1, R Boekenkamp, R Kaulitz, C Fink, G Hausdorf.   

Abstract

A transcatheter technique is described for stabilization and retrieval of an embolized Amplatzer device, which was inadvertently deployed in the left atrium of a 2-year-old girl with hemodynamically significant atrial septal defect within the fossa ovalis. Since surgery was not available for immediate device retrieval, transcatheter means were chosen to prevent embolization of the device into the mitral valve. After stabilizing the device by creating a guidewire circuit through the wire mesh of the right atrial disk, the Amplatzer device was repositioned through the interatrial defect by snaring the microscrew. No residual shunting and perfect device position was recorded during Doppler echocardiography directly after the procedure as well as 1, 2, and 120 days after the intervention. Cathet. Cardiovasc. Intervent. 51:297-300, 2000. Copyright 2000 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2000        PMID: 11066110     DOI: 10.1002/1522-726x(200011)51:3<297::aid-ccd11>3.0.co;2-o

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  5 in total

1.  Successful retrieval of migrated Amplatzer septal occluder.

Authors:  Salil Shirodkar; Sushant Patil; Robin Pinto; Bharat Dalvi
Journal:  Ann Pediatr Cardiol       Date:  2010-01

2.  Transcatheter retrieval of embolized AMPLATZER Septal Occluder.

Authors:  Pravin K Goel; Aditya Kapoor; Aditya Batra; Roopali Khanna
Journal:  Tex Heart Inst J       Date:  2012

3.  Tough, but not impossible - Retrieval of large atrial septal occluder devices embolized to left atrium.

Authors:  Rangaraj Ramalingam; Shivanand Patil; Natraj Setty; Jayashree Kharge; Beeresha Puttegowda; Jayasheelan Mambally Rachaiah; Cholenahally Nanjappa Manjunath
Journal:  Interv Med Appl Sci       Date:  2017-03

4.  Retrieval and repositioning of an embolized atrial septal defect closure device using a gooseneck snare.

Authors:  Akash Batta; Sanjeev Naganur; Ajay Rajan; Kunwer Abhishek Ary; Atit Gawalkar; Parag Barwad
Journal:  Egypt Heart J       Date:  2021-06-05

5.  A novel snare assistance safeguards against early embolization of devices and facilitates quick retrieval of malpositioned devices in atrial septal defects with deficient margins.

Authors:  Sreeja Pavithran; Kothandam Sivakumar
Journal:  Ann Pediatr Cardiol       Date:  2015 Sep-Dec
  5 in total