Dual chamber arrhythmia detection in the implantable cardioverter defibrillator.

B Dijkman1, H J Wellens.   

Abstract

INTRODUCTION: Dual chamber implantable cardioverter defibrillator (ICD) technology extended ICD therapy to more than termination of hemodynamically unstable ventricular tachyarrhythmias. It created the basis for dual chamber arrhythmia management in which dependable detection is important for treatment and prevention of both ventricular and atrial arrhythmias. METHODS AND
RESULTS: Dual chamber detection algorithms were investigated in two Medtronic dual chamber ICDs: the 7250 Jewel AF (33 patients) and the 7271 Gem DR (31 patients). Both ICDs use the same PR Logic algorithm to interpret tachycardia as ventricular tachycardia (VT), supraventricular tachycardia (SVT), or dual (VT+ SVT). The accuracy of dual chamber detection was studied in 310 of 1,367 spontaneously occurring tachycardias in which rate criterion only was not sufficient for arrhythmia diagnosis. In 78 episodes there was a double tachycardia, in 223 episodes SVT was detected in the VT or ventricular fibrillation zone, and in 9 episodes arrhythmia was detected outside the boundaries of the PR Logic functioning. In 100% of double tachycardias the VT was correctly diagnosed and received priority treatment. SVT was seen in 59 (19%) episodes diagnosed as VT. The causes of inappropriate detection were (1) algorithm failure (inability to fulfill the PR<RP condition in atrial tachyarrhythmias with 1:1 AV conduction, and far-field R wave sensing intermittently present during sinus tachycardia); (2) programming settings (atrial fibrillation/atrial flutter with ventricular rate above the SVT limit); and (3) algorithm limitations (atrial tachycardia with ventricular rate around the shortest programmable SVT limit and SVT redetection following VT therapy). Programming measures improved detection ability in 13 of 59 of inappropriately detected arrhythmias.
CONCLUSION: Dual chamber detection algorithms evaluated in a subset of diagnostically difficult arrhythmias allow safe detection of double tachycardias but require further extension and programmability to improve VT:SVT discrimination rules.

Entities:  

Mesh:

Year:  2000        PMID: 11059974     DOI: 10.1111/j.1540-8167.2000.tb01756.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  3 in total

1.  Single-brand dual-chamber discriminators to prevent inappropriate shocks in patients implanted with prophylactic implantable cardioverter defibrillators: a propensity-weighted comparison of single- and dual-chamber devices.

Authors:  Sem Briongos-Figuero; Ana Sánchez; M Luisa Pérez; José B Martínez-Ferrer; Enrique García; Xavier Viñolas; Ángel Arenal; Javier Alzueta; Nuria Basterra; Aníbal Rodríguez; Ignacio Lozano; Roberto Muñoz-Aguilera
Journal:  J Interv Card Electrophysiol       Date:  2018-12-06       Impact factor: 1.900

2.  A trial design for evaluation of empiric programming of implantable cardioverter defibrillators to improve patient management.

Authors:  John M Morgan; Laurence D Sterns; Jodi L Hanson; Kevin T Ousdigian; Mary F Otterness; Bruce L Wilkoff
Journal:  Curr Control Trials Cardiovasc Med       Date:  2004-11-12

Review 3.  ICD discrimination of SVT versus VT with 1:1 V-A conduction: A review of the literature.

Authors:  Rhanderson N Cardoso; Chris Healy; Juan Viles-Gonzalez; James O Coffey
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.