| Literature DB >> 11057508 |
P Rohner1, B Ninet, A M Benri, R Auckenthaler.
Abstract
The Bactec MGIT 960 system (Becton Dickinson, USA), designed for the culture of mycobacteria, was compared with the Bactec 460 instrument and culture on two solid egg-based media using a total of 1024 clinical specimens. Mycobacteria could be identified from 99 (9.7%) specimens, 89 (90%) of which were identified by the Bactec 960 system, 90 (91%) by the Bactec 460 system, and 82 (83%) by culture on the two egg-based media. The Bactec 960 cultures became positive an average of 16.7 days after specimen collection, the Bactec 460 cultures 14.9 days after collection, and the cultures on egg-based media 26.2 days after collection. The Bactec 960 is a compact and highly automated nonradiometric system that may replace the Bactec 460 system.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11057508 DOI: 10.1007/s100960000344
Source DB: PubMed Journal: Eur J Clin Microbiol Infect Dis ISSN: 0934-9723 Impact factor: 3.267