BACKGROUND: Adequate humidification in long-term jet ventilation is a critical aspect in terms of clinical safety. AIM: To assess a prototype of an electronic jet-ventilator and its humidification system. METHODS:Forty patients with respiratory insufficiency were randomly allocated to one of four groups. The criterion for inclusion in this study was respiratory insufficiency exhibiting a Murray score above 2. The four groups of patients were ventilated with three different respirators and four different humidification systems. Patients in groups A and B received superimposed high-frequency jet ventilation (SHFJV) by an electronic jet-ventilator either with (group A) or without (group B) an additional humidification system. Patients in group C received high-frequency percussive ventilation (HFPV) by a pneumatic high-frequency respirator, using a hot water humidifier for warming and moistening the inspiration gas. Patients in group D received conventional mechanical ventilation using a standard intensive care unit respirator with a standard humidification system. SHFJV and HFPV were used for a period of 100 h (4days). RESULTS: A significantly low inspiration gas temperature was noted in patients in group B, initially (27.2 +/- 2.5 degrees C) and after 2 days (28.0 +/- 1.6 degrees C) (P < 0.05). The percentage of relative humidity of the inspiration gas in patients in group B was also initially significantly low (69.8 +/- 4.1%; P < 0.05) but rose to an average of 98 +/- 2.8% after 2 h. The average percentage across all four groups amounted to 98 +/- 0.4% after 2 h. Inflammation of the tracheal mucosa was found in patients in group B and the mucosal injury score (MIS) was significantly higher than in all the other groups. Patients in groups A, C and D showed no severe evidence of airway damage, exhibiting adequate values of relative humidity and temperature of the inspired gas. CONCLUSION: The problems of humidification associated with jet ventilation can be fully prevented by using this new jet-ventilator. These data were sustained by nondeteriorating MIS values at the end of the 4-day study period in groups A, C and D.
RCT Entities:
BACKGROUND: Adequate humidification in long-term jet ventilation is a critical aspect in terms of clinical safety. AIM: To assess a prototype of an electronic jet-ventilator and its humidification system. METHODS: Forty patients with respiratory insufficiency were randomly allocated to one of four groups. The criterion for inclusion in this study was respiratory insufficiency exhibiting a Murray score above 2. The four groups of patients were ventilated with three different respirators and four different humidification systems. Patients in groups A and B received superimposed high-frequency jet ventilation (SHFJV) by an electronic jet-ventilator either with (group A) or without (group B) an additional humidification system. Patients in group C received high-frequency percussive ventilation (HFPV) by a pneumatic high-frequency respirator, using a hot water humidifier for warming and moistening the inspiration gas. Patients in group D received conventional mechanical ventilation using a standard intensive care unit respirator with a standard humidification system. SHFJV and HFPV were used for a period of 100 h (4days). RESULTS: A significantly low inspiration gas temperature was noted in patients in group B, initially (27.2 +/- 2.5 degrees C) and after 2 days (28.0 +/- 1.6 degrees C) (P < 0.05). The percentage of relative humidity of the inspiration gas in patients in group B was also initially significantly low (69.8 +/- 4.1%; P < 0.05) but rose to an average of 98 +/- 2.8% after 2 h. The average percentage across all four groups amounted to 98 +/- 0.4% after 2 h. Inflammation of the tracheal mucosa was found in patients in group B and the mucosal injury score (MIS) was significantly higher than in all the other groups. Patients in groups A, C and D showed no severe evidence of airway damage, exhibiting adequate values of relative humidity and temperature of the inspired gas. CONCLUSION: The problems of humidification associated with jet ventilation can be fully prevented by using this new jet-ventilator. These data were sustained by nondeteriorating MIS values at the end of the 4-day study period in groups A, C and D.
High-frequency jet ventilation (HFJV) has proved to be an alternative
to conventional mechanical ventilation (CMV) [1,2,3,4]. The major advantage of HFJV lies in the improvement of the
mucociliary transport system [5], the recruitment of
atelectatic areas [6], and the improvement of oxygenation
[7,8], maintaining very low tidal
volumes to avoid lung barotrauma [9]. However, one of the
critical issues concerning HFJV is the adequate humidification and warming of
the inspired gases [10]. Specific problems derive from
the physical phenomena of the high velocity of the jet-stream, the
Joule-Thompsen effect and the Venturi effect [11],
causing low temperature and low relative humidity of the inspired gases [12,13].There are several forms of HFJV [14]. Combined
high-frequency jet ventilation (CHFJV) is a technique that requires a
conventional respirator (endotracheal tube) and a high-frequency jet-ventilator
(endotracheal jet tube) [15]. A special type of the
CHFJV is the superimposed high-frequency jet ventilation (SHFJV), which is a
time-regulated, pressure-controlled ventilation technique. It is characterized
by the simultaneous application of a low-frequency jet stream and a
high-frequency jet stream, resulting in the superposition of two jet streams
with different frequencies [16]. This ventilation
technique is managed by one respirator. Volumetric diffusive respiration is a
type of ventilation (using, for example, the VDR 4, Percussionaire Corp, Idaho,
USA) named high-frequency percussive ventilation (HFPV) [17]. Although these techniques have been described in the
literature, they have not been rigorously compared to each other in long-term
applications. There are several humidification systems for CMV [18] and HFJV [19,20], for example, hot water humidifiers, cold water
humidifiers and heat and moisture exchangers. The most commonly used
humidification system in our institution is the hot water humidifier (Aquapor,
Type 8406640, Draeger Corp, Luebeck, Germany).The aim of this study was to show that the problems of humidification
associated with SHFJV can be prevented by using the correct humidification
system. Proper methods for showing possible epithelial damage were used.
Material and methods
After approval by the institutional Ethics Committee and informed
consent, 40 intensive care unit (ICU) patients (Table 1)
were randomly allocated to one of four groups (A, B, C or D; 10 patients in
each group). The inclusion criterion was respiratory insufficiency due to
pneumonia or involvement of the lungs in multi-organ dysfunction syndrome. All
patients showed a Murray score [21] (Table
2) above 2.0, fulfilling the criteria for acute
respiratory distress syndrome (ARDS) or acute lung insufficiency (ALI)
according to the European American Consensus Conference 1994 (Table
2). The mean number of CMV respiration days before group
allocation was 2.2 days. The median age of the 16 female and 24 male patients
was 67 years, ranging from 55 to 79 years.
Table 1
Patient characteristics and demographic data at study entry
Characteristic
Age (years)
67 ± 12
Sex (male/female)
24/16
Etiology of acute respiratory failure (n)
Postoperative
20
Pneumonia
12
Sepsis
8
Drug treatment
Midazolam (n, mg/kg per h)
40, 0.12 ± 0.03
Sufentanil (n, mg/kg per h)
40, 0.012 ± 0.003
Ketamine (n, mg/kg per h)
15, 1.5 ± 0.5
Respirator therapy (CPPV, pressure controlled mode, by
Evita*)
All patients were sedated with midazolam 0.1–0.15 mg/kg per h and
sufentanil 0.01–0.015 mg/kg per h, and optionally with ketamine 1–2 mg/kg per h.
Fluid and catecholamines were administered to achieve a pulmonary capillary
wedge pressure (PCWP) between 12 and 18 mmHg, a cardiac index (CI) above
2.5 l/min per m2, and a systemic vascular resistance index (SVRI)
between 600 and 1500 dyn × s/cm5 per m2. Antibiotic
therapy was adapted according to the results of bacterial cultures of the
tracheobronchial secretion. Red packed cells where given to restore blood
volume when required and to maintain the hematocrit >32%. Hemodynamic
variables were assessed through central venous and peripheral arterial lines.
In eight patients, a Swan-Ganz pulmonary artery catheter was inserted.Patients in group A received SHFJV by a prototype of an electronic
jet-ventilator (Alexander 1, Festo Corp, Vienna, Austria) (Fig. 1). The Alexander 1 works in an open system, providing a
low-frequency jet stream ranging from 4 to 40 breaths/min that influences the
peak pressure, and a high-frequency jet stream ranging from 60 to 990
breaths/min that influences the positive end-expiratory pressure (PEEP). Both
jet streams are simultaneously applied to the patient using a newly developed
jet adapter [22]. This plastic jet adapter (T-connector
with a luer lock, Willy RãSCH AG, Kermen, Germany) consists of a T-piece
and four central small-bore cannulas that can be connected to any commercially
available endotracheal tube, avoiding reintubation with an endotracheal jet
tube when beginning SHFJV. The high-frequency and low-frequency ventilation is
performed with two jet nozzles which have been designed according to optimal
dynamic flow measurements. Two further jet nozzles are used for continuous
airway pressure monitoring and for passing saline solution into the jet stream.
Next to the T-piece there is a lockable aperture that can be used for
suctioning or bronchoscopy, without loss of PEEP.
Figure 1
A prototype of an electronic jet ventilator (Alexander 1; Festo
Corp, Vienna, Austria). This ventilator delivers a high-frequency and a
low-frequency jet stream simultaneously to the patient: superimposed
high-frequency jet ventilation (SHFJV). PEEP, positive end-expiratory pressure;
FiO2, fractional inspiratory oxygen concentration; LF, lowfrequency;
HF, high frequency.
For humidification, we used a combined humidification and warming
system for patients in group A. The entrained gases (bias flow) were humidified
by a hot water humidifier (Aquapor) and the inspiration gas bubbles through a
waterbath heated by an immersed heating element. Gas leaving the device is
saturated with water vapor and is heated to a pre-set temperature. The absolute
humidity can be altered by changing the temperature of the waterbath.Humidification of the jet gas was achieved by a continuous infusion of
0.9% saline via a separate cannulla in the jet adapter. This was propelled and
nebulized by the high-pressure jet stream, starting at 20 ml/h. The saline
solution was warmed to 39°C by a fluid warmer (HL-90 INT, Level 1
Technologies Corp, Rockland, Massachusetts, USA) before it reached the jet
adapter (ie, an additional humidification system). This allows for a possible
warm-up of 42°C to compensate for the temperature drop of the gas after
decompression (the Joule-Thompsen effect).Patients in group B also used the Alexander 1, with a hot water
humidifier for the bias flow (Aquapor) and a continuous saline infusion for the
jet flow (the additional fluid warmer mentioned above was not used in patients
in group B).Patients in group C received HFPV using the VDR 4, which is a
pneumatic time-regulated, pressure-controlled respirator that works in a closed
system to provide two oscillating pressure plateaus. The HFPV is generated by a
phasitron, causing gases in the airways to oscillate back and forth between the
respirator and the patient, with the fresh gas entrained from the bias flow.
Because of the continuous flow applied during HFPV the patient is able to
breathe spontaneously at any time during the respiratory cycle, and respiratory
weaning can be performed. This method allows conventional heater humidifiers to
be used. In our study we used a hot water humidifier (Aquapor).Patients in group D received CMV using a conventional intensive care
respirator (Evita, Draeger Corp) in a pressure-controlled mode using a hot
water humidifier (Aquapor).In all four groups, PEEP, the fractional inspiratory oxygen
concentration (FiO2) and inspiration to expiration time ratio (I:E)
were adjusted to keep arterial oxygen saturation
(SaO2) above 90% with the lowest possible peak airway
pressure and FiO2.On the first day of the study, relative humidity measurements were
performed at jet ventilation commencement, and after 20 min and 2 h.
Consequently, data were taken only twice a day until the end of the 4-day
study. Values were accurate to ± 2% of relative humidity. Relative
humidity was measured inside the tube, at its distal end, with an electronic
device (E20-FXD, E u. E Elektronik Corp, Unterwiesen, Austria). This very small
sensor consists of a condensator with a capacity of 600 picoFarad at a relative
humidity of 70%. The polymer sensor changes its capacity according to the
amount of moisture inside the area that is measured. This chemical reaction is
translated into an electronic signal, which is shown on a liquid crystal
display (ie, monitor) outside the tube as a percentage of the relative humidity
inside the inspired gas flow.The temperature was measured at the same location following the same
timetable as above with a temperature sensor line (21076 A, Hewlett Packard,
Palo Alto, California, USA). Values were accurate to ± 0.1°C.The tracheobronchial mucosa was inspected twice a day by bronchoscopy
(Olympus BF Type P20D, Olympus Corp Ltd, Shirakwa, Japan) over the entire 100-h
study period according to a mucosal injury score (MIS). The MIS takes into
account the area and the extent of macroscopic epithelial damage in the
tracheobronchial system: area A represents the mucosa around the cuff (for
inspection, the tube has to be withdrawn a little); area B represents the
mucosa from below the cuff down to the carina; area C represents the mucosa of
the right mainstem bronchus; and area D represents the mucosa of the left
mainstem bronchus. Each area is given a number, representing the amount of
mucosal damage (0, no exsiccosis, no inflammation and no necrosis; 1,
exsiccosis as shown by a pale mucosa with no signs of epithelial damage; 2,
inflammation of the mucosa shown by mucosal redness and tumor; 3, necrosis
shown by a pale mucosa with signs of epithelial damage). Thus, the possible
value of the MIS ranges from 0 to 12 with damage classified as: 1–3, simple
damage; 4–8, significant damage; 9–12, severe damage.The humidification rate (initially 20 ml/h) was set depending on the
bronchoscopic aspect of the tracheobronchial mucosa; classification above
'simple damage' (ie, MIS above 3) led to an increase in the
humidification rate up to and above 45 ml/h.
Statistical analysis
For demographic data (Table 1) and measured
values of relative humidity and gas temperature a Duncan test was calculted to
determine statistical significance. All values are expressed as mean ±
standard deviation. P < 0.05 was regarded as statistically
significant.Since homogeneity of variances were seen, statistical evaluation was
performed using single factorial analysis. However, as the small number of spot
samples impaired the standard distribution, a nonparametric method (variance
analysis by Kruskal–Wallis) was also used. Both methods showed the same
significant differences between the groups. Furthermore, a pathanalysis was
calculated to check the statistical relevance of the hypothetical influence of
the forms of ventilation on temperature and humidity.A prototype of an electronic jet ventilator (Alexander 1; Festo
Corp, Vienna, Austria). This ventilator delivers a high-frequency and a
low-frequency jet stream simultaneously to the patient: superimposed
high-frequency jet ventilation (SHFJV). PEEP, positive end-expiratory pressure;
FiO2, fractional inspiratory oxygen concentration; LF, lowfrequency;
HF, high frequency.Patient characteristics and demographic data at study entry*Draeger Corp, Luebeck, Germany. CPPV, continuous
positive pressure ventilation; Pmax, maximal airway pressure; PEEP, positive
end-expiratory pressure; FiO2, fractional inspiratory oxygen
concentration; I:E, inspiration to expiration time ratio.Respiratory insufficiency as the inclusion criterion (Murray score
[21] and European American Consensus Conference 1994
values)PaO2, partial pressure of oxygen; FiO2,
fractional inspiratory oxygen Concentration; PEEP, positive end-expiratory
pressure; ARDS, acute respiratory distress syndrome; ALI, acute lung
insufficiancy; ap, anterior-posterior; PCWP, pulmonary capillary wedge
pressure.
Results
Prior to random allocation to the four groups, all patients were
conventionally ventilated (CMV using Evita) for 48–72 h. The average number of
ventilation days in each group was 10 days (the duration of mechanical
ventilation before weaning started). The overall period of ventilation was
18 ± 4 days.
Relative humidity
Initially, the lowest percentage of mean relative humidity (Fig.
2) was seen in patients in group B (69.8 ± 4.1%;
P < 0.05). Patients in group A showed a mean relative humidity of
71.2%, patients in group C showed a mean relative humidity of 92.2% and
patients in group D showed a mean relative humidity of 92.0%.
Figure 2
Percentage of relative humidity (RH) of the inspiration gas for
each group initially, after 20 min, 2 h, after 2 days and after 4 days of
ventilation (mean± standard deviation).
After 20 min of ventilation the mean relative humidity of the
inspiration gas was still lower in patients in group B (92.8%) compared with
patients in group A (96.8%), group C (97.4%), and group D (98.6%).After 2 h of ventilation patients in all four groups showed almost
equivalent mean values (98 ± 0.4%). All measurements taken from this point
until the end of the 100-h study protocol showed no more significant changes
compared to the values measured after 2 h.
Temperature
Initially, the lowest temperature measured inside the tube (Fig.
3) was seen in patients in group B (27.2 ±
2.5°C; P < 0.05). Patients in group A had a mean gas
temperature of 31.4 ± 2.8°C, those in group C had a mean gas
temperature of 32.1 ± 2.6°C, and those in group D had a mean gas
temperature of 34.2 ± 2.7°C.
Figure 3
Temperature of the inspiration gas of each group initially, after
20 min, 2 h, after 2 days and after 4 days of ventilation (mean ± standard
deviation).
After 20 min of ventilation the inspiration gas temperature was still
lower in patients in group B (27.1 ± 1.8°C) compared with patients in
group A (32 ± 1.8°C), patients in group C (32.5 ± 2.3°C), and
patients in group D (34.1 ± 2.5°C).After 2 h of ventilation the trend was similar (patients in group B,
27.1 ± 1.8°C; patients in group A, 32.4 ± 1.1°C; patients in
group C, 32.6 ± 1.6°C; and patients in group D, 34.3 ±
2.3°C).After 2 days the values were: patients in group B, 28.0 ±
1.6°C; patients in group A, 33.2 ± 1.7°C; patients in group C,
33.2 ± 2.5°C; patients in group D, 34.5 ± 1.8°C.After 4 days the values were: patients in group B, 28.0 ±
1.9°C; patients in group A, 33.0 ± 1.7°C; patients in group C,
33.6 ± 2.5°C; patients in group D, 34.3± 1.8°C.
Mucosal injury score
During the course of SHFJV, patients in group B showed signs of
epithelial lesions and inflammation of the mucosa in area B within hours
(MIS = 2), deteriorating on the third and fourth day (MIS = 3) (Fig. 4). During bronchoscopy, a tracheobronchial secretion was
removed. The bronchial epithelium (area C) showed no pathologic evidence.
Typical changes to the mucosa were not detected in patients in any group other
than those in group B.
Figure 4
Mucosal injury score of each group initially, after 20 min, 2 h,
after 2 days and after 4 days of ventilation (mean± standard
deviation).
In group A, C and D no epithelial damage could be found in any area.
Eight patients (group independent) who were in an intermittent prone position
(usually turned every 12 h) showed little tracheal damage at the end of the
tube.Although one patient in group C died as a result of severe sepsis,
their tracheobronchial mucosa showed no adverse tissue change when compared to
the control group. As the period of ventilation (72 h) was probably sufficient
to induce any injury as a result of the humidification system, data from this
patient were not excluded in our comparison. The statistical relevance of the
hypothetical influence of the forms of ventilation on temperature and humidity
were checked by pathanalysis. Only group B appeared to show any influence.
Humidification rate of the additional humidification system
Initially, the humidification rate of the additional humidification
system was set at 20 ml/h saline solution via the jet adapter (Table
3). This was changed depending on the bronchoscopic
aspect of the tracheobronchial mucosa and the presence of dry secretion.
Patients in group A showed an average setting of 30 ± 10 ml/h, starting at
20 ml/h and rising to 40ml/h on the second day. Patients in group B showed the
highest demand for saline solution, with an average setting of 45 ± 5 ml/h
and a peak of 50 ml/h on the fourth day. Patients in groups C and D were served
by a hot water humidifier without an additional humidification system.
Table 3
Mucosal injury score and humidification rate settings during the
course of superimposed high-frequency jet ventilation
Group A
Group B
Initially
Mucosal injury score
0
0
Humidification rate setting (ml/h)
20
20
After 20 min
Mucosal injury score
0
1
Humidification rate setting (ml/h)
20
30
After 2 h
Mucosal injury score
1
2
Humidification rate setting (ml/h)
30
40
After 2 days
Mucosal injury score
2
3
Humidification rate setting (ml/h)
40
45
After 4 days
;Mucosal injury score
1
3
Humidification rate setting (ml/h)
35
50
Percentage of relative humidity (RH) of the inspiration gas for
each group initially, after 20 min, 2 h, after 2 days and after 4 days of
ventilation (mean± standard deviation).Temperature of the inspiration gas of each group initially, after
20 min, 2 h, after 2 days and after 4 days of ventilation (mean ± standard
deviation).Mucosal injury score of each group initially, after 20 min, 2 h,
after 2 days and after 4 days of ventilation (mean± standard
deviation).Mucosal injury score and humidification rate settings during the
course of superimposed high-frequency jet ventilation
Discussion
This study yielded two major findings. Firstly, adequate
humidification without sufficient warming of the inspiration gas does not
prevent epithelial damage or inflammation. Secondly, the amount of
humidification needed has to be changed almost every day, sometimes even within
hours. Thus, regular bronchoscopy is necessary. Leaving the inspiration gas
temperature significantly below 30°C ('cold ventilation'),
even under high humidity (relative humidity above 90%), leads to epithelial
damage, inflammation or even necrotizing tracheobronchitis [23].The continuous 0.9% saline infusion into the humidification line of
the jet adapter started at 20 ml/h. Regular checks on the mucosa showed that
over 75% of all patients needed higher humidification over the whole study
period, in some cases an increase of 200%. This increase was dependent on the
bronchoscopic aspect of the tracheobronchial mucosa and the presence of dry
secretions. Detecting these changes for alteration of the humidification
settings requires a lot of experience and cannot be explained merely by facts,
figures and equations.To humidify a dry gas and reach 100% relative humidity, however,
requires 44mg water to be added per litre of inspired gas. As a consequence,
the rate of humidification (QH2O; ml/h) equals 2.64 × V, where V
is the minute ventilation. Because minute ventilation during high-frequency
ventilation ranges between 20 and 40 l/min, QH2O may vary between 50
and 100 ml/h. Of course, any change in the ventilatory parameters (tidal volume,
ventilation frequency per min, driving pressure, I:E) modifies minute
ventilation and, therefore, QH2O. These basic physical principles of
humidification are well known, and are accepted in ventilation therapy. The
saline solution was warmed by a fluid warmer to 39°C before it reached the
jet adapter, allowing for a possible warm up of 42°C if necessary, to
compensate for the temperature drop of the gas after decompression
(Joule-Thompsen effect) and possible epithelial lesions.In fact, increasing the temperature of the inspired gas delivered
during high-frequency jet ventilation from 39°C–42°C might appear
insufficient. After decompression of the gas into the trachea, there is a
sudden drop in the temperature of 5–10°C. As a consequence, to reach
37°C in the tracheobronchial tree, inspired gas should be warmed to at
least 45°C.In the face of this argument, our humidification system seems to face
a serious drawback by providing insufficient temperature of the inspired gas,
especially in terms of inducing lesions in the tracheobronchial tree after a
prolonged period of SHFJV. However, our period of observation covered
approximately 4 days and no signs of persisting inflammation or even necrotizing
tracheobronchitis were observed in any group; only tracheobronchial damage
(epithelial lesions and transitory inflammation) were detected. These
alterations could be compensated for by increasing the humidification rate
(Table 3) of the additional humidification system, which
differed more than 50% from patient to patient.The etiology of these different demands can be seen in the
multifactorial reasons that influence it, the most important of which is the
drift of the so-called isothermic saturation boundary (ISB). This boundary
usually lies above the carina during spontaneous breathing and represents the
point where the inspired air is fully saturated with vapor and has a
temperature of 36°C. Turbulent flow also changes to laminar flow below
this point [24], and intubation and ventilation bring
this ISB further back along the tracheobronchial tract. Thus, the ability of
the lungs to facilitate an adequate temperature and amount of moisture in the
inspiration gas is dramatically impaired, leading to the problems that are
associated with inadequate humidification and warming.At present, equipment to monitor humidity is not sufficiently
sophisticated to allow accurate breath to breath measurements of humidity
within the airway. The estimation of humidification requirements must therefore
be based on scientific evidence and clinical impression. Humidification of
inspired gases should not be considered in isolation but as part of total
airway management. It should be associated with careful fluid balance,
physiotherapy, bronchial aspiration and appropriate drug therapy.Functional studies [24,25] have shown adverse effects of dry inspiratory gas on the
tracheal mucosa during ventilation. Chalon et al [12] studied the cytology of epithelial cells from tracheal
washings taken during anesthesia via an endotracheal tube with gases at
different humidities. Abnormal cytology was found within 2h of ventilation with
dry gas but not at 60% humidity or higher. Doyle et al [18] demonstrated that extensive epithelial damage to the
trachea with destruction of the cilia, tissue inflammation and necrosis
occurred after 72h of CMV with dry gas, but no damage was observed at high
humidity. Other studies concerning jet ventilation [26,27] demonstrated epithelial damage
at the level of the carina and left and right main-stem bronchus. There are
multiple effects of passing warm, dry gas over the tracheal mucosa. Initially,
there is an increase in blood supply to this region [28], probably as a result of release of local vasodilators
[29]. Subsequently, there is an increase in the
osmolality of the mucus secretion as the airway mucosa humidifies the gas
before it enters the lungs [30]. Mucus secretions are
under parasympathetic control and increase in volume after vagal nerve
stimulation [31] or acetylcholine treatment [32]. As the mucus secretion becomes increasingly hyperosmolar
and dry, it is trapped and encapsulated beneath the surface. Cellular injury
occurs and neutrophil sequestration is exaggerated due to the increase in blood
supply. In this situation, with aggravating tissue damage, an exsudate is
formed that appears as blisters below the mucosal surface. Because of the
external force from mechanical ventilation or jet ventilation, the encapsulated
mucus penetrates and causes sloughing of the tracheal epithelial cells.The clinical implications of inadequate humidification during
mechanical ventilation or jet ventilation are very important. The use of dry
gas during anesthesia is still common during surgical procedures. Furthermore,
dry oxygen is used during resuscitation of asphyxiated patients and preterm
infants. The use of dry gas during routine respiratory changes may impair
already compromised lungs, especially in infants with chronic obtructive
pulmonary disease and adults with ARDS. Tarnow-Mordi et al [32,33] showed an increase in the
incidence of pneumothorax and chronic lung disease in patients who were
ventilated with inspired gas at low temperatures, implying a low inspired
absolute humidity. Necrotizing tracheobronchitis [34],
which has been described in several ICUs, has also been attributed to low
inspired humidity leading to sloughing of tracheal and bronchial cells [35]. Buchdahl et al [36]
showed an increase in chronic upper respiratory tract problems in children with
reduced ciliary beat frequency. Damage to the tracheal mucosa occurs during
endotracheal intubation and ventilation whatever the inspired humidity. Using
dry inspired gas the damage is dramatically worse; therefore, clinical
procedures should aim to reduce the use of dry gases in ventilator
circuits.Warming of the inspiration gas is just as important as its
humidification to avoid bronchotracheal damage. A lack of warming of the
inspired gas is said to be responsible for necrotizing tracheobronchitis among
many other cofactors, especially under treatment with HFJV [37,38]. However, Keszler et al
[39] showed in a multicenter clinical trial that the
incidence of necrotizing tracheobronchitis is similar comparing HFJV with CMV
in neonates.Although multiple factors are associated with necrotizing
tracheobronchits, Hanson et al [40] proposed
that regional or generalized airway ischemia was the mechanism for airway
damage, based on the appearance of mucosal damage with a lack of warming.Cavanagh et al [41] found an increase
in regional tracheal blood flow (10.3-fold), measured by radioactive
microspheres, under HFJV compared with spontaneous breathing, increasing
further when using dry gas at 22°C. The authors claim the enhanced local
hyperemia effect is because of greater water and heat extraction. A moderate
hyperventilation that occurred during high-frequency jet ventilation in this
study [41] might also have affected tracheal blood flow,
although the authors concede that the hyperemia effects appear to relate to
shear stress and water removal rather than alveolar partial pressure of carbon
dioxide. They found blood flow improvements in the tracheal mucosa, with the
highest increase using dry gas.Although the study by Cavanagh et al covers only a period of
2.5 h, another study [42] shows that longer periods of
ventilation (33 h) produce no significant differences in airway damage when
comparing CMV with HFJV.Another explanation for the hyperemia effects under HFJV is given by
Baile et al [43]. They claim that greater shear
stresses with HFJV may also alter mucosal epithelia permeability and
secondarily affect the blood flow.Our study showed that, by providing proper humidification and warming
of the inspiration gases, epithelial damage to the tracheobronchial mucosa can
be prevented, as can possible inflammation and necrotizing tracheobronchitis,
even in long-term applications.The damage to the mucosa observed in patients in group B relates well
with the outcome of the studies by Rouby et al [7,8] and Chatburn and McClellan [19], who have made a very interesting contribution to the
problems associated with humidification and warming of high-frequency jet
ventilation. Doyle et al [18] and Chatburn and
McClellan [19] showed that, by providing adequate
moisture and proper temperature of the inspiration gases, deleterious effects
on the tracheobronchial mucosa can be prevented. Although their mechanical
expenditures and material costs were much higher than ours, the outcome of
their studies were just as satisfactory as the results of this study. SHFJV has
been shown to be a serious alternative to CMV. Using the Alexander 1, the
problems of humidification and warming of the ventilation gas can be handled
very well.