BACKGROUND: Multilumen catheters are commonly used in critically ill children. Their use, however, is associated with significant morbidity. We studied the simultaneous administration of incompatible drugs using a new triple-lumen catheter with decreased length and port to port distances. METHODS: Ten domestic swine, 10-20 kg in weight, were divided into two groups of five. Total parenteral nutrition was administered through the distal port and phenytoin was administered as a bolus and as an infusion in each group. Samples were taken from two sites during the bolus and at 1, 5, and 15 min during phenytoin infusion. Histograms were generated for particle size and concentration. Samples were also examined under the microscope for particles. RESULTS: Histograms of particle size did not show any alteration of the histogram that would suggest particle size > 2 µm in diameter in the study or control samples. No particles were identified by phase microscope, light microscope, or Wright stain smear. CONCLUSIONS: The use of a triple-lumen catheter with a distance of 0.4cm between the proximal port and the medial port and 1.3 cm between the medial port and the distal port, for the in vivo simultaneous administration of incompatible solutions does not result in precipitates large enough to cause adverse clinical effects.
BACKGROUND: Multilumen catheters are commonly used in critically ill children. Their use, however, is associated with significant morbidity. We studied the simultaneous administration of incompatible drugs using a new triple-lumen catheter with decreased length and port to port distances. METHODS: Ten domestic swine, 10-20 kg in weight, were divided into two groups of five. Total parenteral nutrition was administered through the distal port and phenytoin was administered as a bolus and as an infusion in each group. Samples were taken from two sites during the bolus and at 1, 5, and 15 min during phenytoin infusion. Histograms were generated for particle size and concentration. Samples were also examined under the microscope for particles. RESULTS: Histograms of particle size did not show any alteration of the histogram that would suggest particle size > 2 µm in diameter in the study or control samples. No particles were identified by phase microscope, light microscope, or Wright stain smear. CONCLUSIONS: The use of a triple-lumen catheter with a distance of 0.4cm between the proximal port and the medial port and 1.3 cm between the medial port and the distal port, for the in vivo simultaneous administration of incompatible solutions does not result in precipitates large enough to cause adverse clinical effects.
A multilumen central venous catheter is the preferred vascular access
route for critically ill patients requiring multiple drug infusions, parenteral
hyperalimentation, and other potentially incompatible drugs [1]. The complexity of delivering these substances becomes more
difficult when the size and length of the catheter is limited, as in the
pediatric population. A previous study has shown that in vivo
simultaneous intravenous infusion of physically incompatible substances through
a commercially available multiple lumen intravenous catheter, double-lumen
peripheral venous catheter (IV–01100, Arrow International, Reading,
Pennsylvania, USA) did not cause precipitation in the vascular system or other
adverse clinical effects [2].This study looked at a modified 5.5 F × 5 cm, triple-lumen catheter
with a port distance of 0.4 cm between the medial and proximal ports and 1.3 cm
between the distal and medial ports (Arrow International), with a total
distance of 1.7 cm between the distal port and the proximal port, to assess if
decreased port spacing between lumens causes precipitation when incompatible
intravenous solutions are administered simultaneously.
Materials and methods
Ten domestic swine weighing 10–20 kg were anesthetized with ketamine 10 mg/kg intramuscularly, atropine 0.01 mg/kg intravenously, and isoflurane for
continuous anesthesia. The trachea was intubated with a #5 or #6 uncuffed
endotracheal tube and controlled ventilation was instituted to maintain normal
blood gases. A lead II electrocardiogram was monitored continuously along with
temperature, respiration, and blood pressure. A peripheral vein was isolated
and cannulated for fluid and anesthetic administration as necessary during the
experiment. The femoral veins in both hind limbs were isolated by the cut-down
technique. The experimental triple-lumen catheter, with decreased port to port
distances, was inserted into the right femoral vein, and the control catheter
was inserted into the left femoral vein under direct visualization. The control
catheter was a 5.5 F × 5 cm, triple-lumen catheter with a standard port
distance of 1.6cm between the proximal and medial ports and 1.6 cm between the
medial and distal ports respectively, for a total distance of 3.2 cm between the
distal port and the proximal port (Arrow International). The size of the lumens
is 20 gauge for the distal lumen, and 22 gauge for the medial and proximal
lumens. Placement was confirmed by aspiration of blood from all ports.A midline incision was made, the peritoneum was identified, and the
incision was extended to isolate and retract peritoneal contents exposing the
retroperitoneal vasculature. The iliac vessels and the inferior vena cava were
dissected and isolated. Sampling sites were identified immediately distal to
the catheter tip and one catheter length distal to the first sampling site. An
18-gauge catheter with a three-way stop-cock was inserted at each sampling site
(Fig 1). Hematocrit was between 35% and 45% during the
experiment.
Phenytoin and total parenteral nutrition (TPN) were shown to be
incompatible in vitro in a previous study [3]. A
solution of phenytoin with a concentration of 50 mg/ml and pH12 as evaluated on
a pH meter (255 Ph/ISE, CIBA Corning Diagnostics, Oberlin, Ohio, USA) was
administered to all animals (n = 10) through the distal lumen at the
usual maintenance dose of 2.5 mg/kg/dose at a rate of 1 mg/kg/min up to a
maximum of 50 mg/min via a syringe infusion pump (1001, Medfusion Systems,
Norcross, Georgia, USA). The bolus was followed by a 3 ml normal saline flush.
The TPN solution (Table 1), with a pH of 5.8, was
administered simultaneously through the medial lumen at the standard clinical
maintenance rate calculated in ml/kg/day and divided over a 24-h period using
the same syringe infusion pump system. In the second phase of the experiment,
the same procedure described above was performed with the phenytoin solution
administered through the medial lumen and the TPN administered through the
distal lumen. In the final phase, the phenytoin was administered through the
proximal lumen and the TPN through the medial lumen. The same methodology was
repeated on the opposite limb using the control catheter.
Table 1
The total parenteral nutrition solution
Amino acids
2.5%
Dextrose
10%
Intralipids
1.25%
NaCl
30 mEq
KCl
10 mEq
Ca gluconate
1000 mEq
K2HPO4
5 mEq
MVI-12
1 ampule
MgSO4
200 mg
Trace elements
5 ml
Total volume
500 ml
Blood sampling was performed mid-way through the phenytoin bolus
administration, and at 1, 5, 10, and 15 min intervals for all ports of both
catheters. Sampling was performed simultaneously at both sampling sites; each
sample consisted of 1 ml whole blood. Histograms were generated by a coulter
counter (Sysmex K-1000, Long Grove, Illinois, USA) for particle size. Quality
control on the analyzer is performed on a real-time basis using commercial
controls once every 8-h shift on a daily basis (Equinox 16T, Hematronix Inc.,
Benicia, California,USA). All quality control data are handled according to
current College of American Pathologists' standards. Calibration of the
analyzer is checked quarterly using a commercial calibrator. Blinded specimens
were examined microscopically by phase microscope, light microscope, and Wright
stain smears.The experimental protocol was reviewed and approved by the animal care
committee of our institution.
Results
Histograms for particle size, set with apertures for white blood cells
(24–300 μ m/100 μ l), red blood cells (30–250 μ m/100 μ l)
and platelets (2–20μ m/100 μ l), did not show any alteration of the
histogram suggesting particle size > 2 μ m in diameter. At no point did
any of the samples, control or study, fail to pass a 2 μ m aperture. Upon
review of the white blood cells, red blood cells, and platelet histograms
generated by the hematology analyzer, we could not identify any foreign
particles in the animal blood. We could not rule out the presence of particles
in very low concentrations, ie ≤ 3 × 103 particles/μ
l, which would be the acceptable background count on this instrument. No
particles from various samples could be identified as examined by phase
microscope, light microscope and Wright stain smears in both control and study
groups.Mean heart rate, respirations, and temperature were 134 ± 8
beat/min, 22 ± 4 breath/min, and 38 ± 0.7°C, respectively.Peripheral multilumen catheter cutdown sites and proximal venous
blood sampling sites.The total parenteral nutrition solution
Discussion
Central venous catheters in the pediatric population, and especially
in the pediatric intensive care setting, are commonly used for the
administration of intravenous fluids, drugs, chemotherapeutic agents, inotropic
agents, and hyperalimentation. Access to the intravascular space includes the
femoral, subclavian, and internal jugular veins. The procedure is not without
complications; these include bleeding, venous thrombosis, vessel perforation,
sepsis, dysrhythmias, and catheter dislodgment or leakage, among others [4,5,6].
Complications are relatively common in the pediatric and neonatal population
with a reported incidence of 11.6% and 53% in two separate studies [7,8]. Leakage and extravasation of drugs
and fluids can have major consequences for the patient. Local edema,
inflammation, infection, and necrosis are the most serious complications of
fluid extravasation and may lead to tissue loss requiring reconstructive
surgery and, in some cases, may cause loss of extremities. Dislodgment of the
catheter can also lead to the death of the patient [9].A plausible reason for this high incidence of fluid extravasation may
be the distance between the proximal and distal ports in commercially available
multilumen pediatric catheters. While the distal lumen may lie within the
intravascular space, the proximal lumen may be close to the site of entry of
the vein or even outside the vein. This increases the possibility of
extravasation if the catheter is accidentally dislodged or is not sutured
properly.The new triple-lumen catheter, with decreased port to port distances
and shorter catheter length compared to commercially available catheters, was
developed for use in the smallest possible patients. The shorter port to port
distance and catheter length may help to minimize the chance that one of the
lumens might be positioned improperly, resulting in the potential for
extravascular fluid infusion or catheter dislodgment.Our findings are consistent with a previous study [2] showing that the simultaneous infusion of phenytoin and TPN
solutions did not cause the precipitation of particles large enough to be of
clinical significance.
Conclusion
Using previously published methodology for the study of the
simultaneous administration of incompatible drugs via a multiple lumen
catheter, we conclude that the use of a triple-lumen catheter with a port
distance of 0.4 cm between the medial and proximal ports, 1.3 cm between the
distal and medial ports, 1.7 cm between the distal and proximal lumens, and
overall length of 5 cm, for the in vivo administration of incompatible
solutions, phenytoin and TPN, using a swine model, did not lead to precipitates
large enough to cause adverse clinical effects in our study. This modified
catheter was developed to minimize the length of the catheter for use in the
smallest possible patients and to decrease the possibility that one of the
lumens might be positioned improperly, resulting in the potential for
extravasation of fluids or drugs.
Authors: M F Goutail-Flaud; M Sfez; A Berg; G Laguenie; C Couturier; F Barbotin-Larrieu; C Saint-Maurice Journal: J Pediatr Surg Date: 1991-06 Impact factor: 2.545
Authors: V Hruszkewycz; P C Holtrop; D G Batton; R S Morden; P Gibson; J D Band Journal: Infect Control Hosp Epidemiol Date: 1991-09 Impact factor: 3.254