S R Meikle1, S Eberl, R R Fulton, M Kassiou, M J Fulham.
Abstract
We investigated the influence of tomograph sensitivity on reliability of parameter estimation in positron emission tomography studies of the rat brain. The kinetics of two tracers in rat striatum and cerebellum were simulated. A typical injected dose of 10 MBq and a reduced dose of 1 MBq were assumed. Kinetic parameters were estimated using a region of interest (ROI) analysis and two pixel-by-pixel analyses. Striatal binding potential was estimated as a function of effective tomograph sensitivity (S(eff)) using a simplified reference tissue model. A S(eff) value of > or =1% was required to ensure reliable parameter estimation for ROI analysis and a S(eff) of 3-6% was required for pixel-by-pixel analysis. We conclude that effective tomograph sensitivity of 3% may be an appropriate design goal for rat brain imaging.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11056379 DOI: 10.1016/s0969-8051(00)00105-0