Initial stability of fully and partially cemented femoral stems.

L Claes1, S Fiedler, M Ohnmacht, G N Duda.   

Abstract

OBJECTIVE: To test the initial stability of a newly designed partially cemented femoral stem in comparison with a fully cemented conventional stem.
DESIGN: An in vitro study to determine the interface motion between femoral stem and bone as a response to loading.
BACKGROUND: The aim of the new prosthesis design is a proximal load transfer by a defined partial cement fixation in the proximal femur region and a slim prosthesis stem in the distal region. Before a clinical study can be started, the new stem has to show an initial stability comparable to that of fully cemented prostheses.
METHOD: Six paired fresh cadaveric femora were used for the testing of the new partially cemented stem (Option 3000, Mathys Orthopaedics, Bettlach, Switzerland) and a fully cemented stem (Weber Shaft, AlloPro, Baar, Swizerland). Under cyclic loading up to 1600 N hip joint forces, the interface motion between implants and bone was measured at six locations.
RESULTS: Both stems showed uncritical interface motions below 43 microm. However, the Option 3000 stem exhibited significantly smaller motions in the proximal region and slightly larger movements in the distal regions than the Weber prosthesis.
CONCLUSIONS: The new type of partially cemented stem provided a comparable initial stability to the fully cemented Weber prosthesis. Relevance The high initial stability of the Option 3000 stem justified the clinical use of the new implant. More than 100 implantations in the last three years, with very good preliminary clinical results, support the preclinical findings.

Mesh:

Substances:

Year:  2000        PMID: 11050357     DOI: 10.1016/s0268-0033(00)00044-9

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  4 in total

1.  Proximal cement fixation in total hip arthroplasty--first results with a new stem design.

Authors:  Thomas Einsiedel; Florian Gebhard; Ilaria Bregolato; Anja Hiemeier; Lothar Kinzl; Markus Schultheiss
Journal:  Int Orthop       Date:  2007-03-13       Impact factor: 3.075

2.  The effect of abductor muscle and anterior-posterior hip contact load simulation on the in-vitro primary stability of a cementless hip stem.

Authors:  Youngbae Park; Carolyne Albert; Yong-San Yoon; Göran Fernlund; Hanspeter Frei; Thomas R Oxland
Journal:  J Orthop Surg Res       Date:  2010-06-24       Impact factor: 2.359

3.  Dynamic mechanical behavior of PMMA based bone cements in wet environment.

Authors:  R De Santis; F Mollica; L Ambrosio; L Nicolais; D Ronca
Journal:  J Mater Sci Mater Med       Date:  2003-07       Impact factor: 3.896

4.  In vivo hip joint loading during post-operative physiotherapeutic exercises.

Authors:  Verena Schwachmeyer; Philipp Damm; Alwina Bender; Jörn Dymke; Friedmar Graichen; Georg Bergmann
Journal:  PLoS One       Date:  2013-10-29       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.