Literature DB >> 11049127

Transesophageal echocardiographic predictors for successful transcatheter closure of defects within the oval fossa using the CardioSEAL septal occlusion device.

T S Momenah1, D B McElhinney, M M Brook, P Moore, N H Silverman.   

Abstract

OBJECTIVES: To define the utility of transesophageal echocardiography in predicting the likelihood of a successful procedure and residual shunting in patients undergoing transcatheter closure of defects within the oval fossa using the CardioSEAL device.
BACKGROUND: Transesophageal echocardiography is used to monitor transcatheter closure of interatrial defects within the the oval fossa, but predictors of successful closure and residual shunting have yet to be determined.
METHODS: We reviewed transesophageal echocardiograms obtained from 26 consecutive patients undergoing attempted transcatheter closure of interatrial defects within the oval fossa between January, 1997 and May, 1998. Assessment of the atrial septum, the septal defect, and the rims of the oval fossa bordering the defect was performed in 3 planes: longitudinal, 4-chamber, and basal short-axis.
RESULTS: Closure proved successful in 24 patients (92%). The defect was significantly larger, and the anterosuperior rim of the defect smaller, in the 2 patients in whom occlusion was not successful. Residual shunting 24 hrs after closure was detected in 14 patients. Significant predictors of leakage included smaller posterior and superior rims, a larger shunt prior to closure, and herniation of a one left atrial arm of the device into the right atrium. In all cases, the sites of leakage were the superior rim of the defect at the superior cavo-atrial junction, and the anterosuperior rim behind the aortic root. Herniation of a left atrial arm into the right atrium was seen in 7 patients (29%). In all, it was the anterosuperior arm which herniated Doppler color flow was suboptimal in detecting residual leaks, and was enhanced substantially with the use of contrast echocardiography.
CONCLUSIONS: Transesophageal echocardiography allows excellent assessment of the oval fossa and deficiencies of its floor in all of their dimensions. It is an important tool for guiding the deployment of the occlusion device in patients undergoing attempted transcatheter closure of defects within the fossa. Contrast echocardiography should be used for optimal detection of residual shunting.

Entities:  

Mesh:

Year:  2000        PMID: 11049127     DOI: 10.1017/s1047951100008209

Source DB:  PubMed          Journal:  Cardiol Young        ISSN: 1047-9511            Impact factor:   1.093