Properties of a hybrid plaster-fibreglass cast.

M N Charles1, D Yen.   

Abstract

OBJECTIVE: To examine the suitability of a plaster-fibreglass hybrid cast for orthopedic applications, comparing them to plaster of Paris (POP) and fibreglass constructs.
METHOD: Groups of 10 standardized hybrid, POP and fibreglass casts were studied. An Instron servohydraulic system was used to test the casts in 3-point bending and shear. OUTCOME MEASURES: Strength, stiffness, weight, thickness and cost of the 3 types of cast, and shear strength at the interface between the POP and fibreglass in the hybrid casts.
RESULTS: The hybrid casts were twice as strong as the POP constructs, were stiffer and weighed 14% less but were thicker and cost 2.5 times more. They were almost as strong as and less than half the cost of the fibreglass constructs but were thicker, not as stiff, and weighed 42% more. The shear strength of the POP-fibreglass interface in the hybrid casts was higher than the 3-point bending strength of this construct by a factor of 3.
CONCLUSIONS: Plaster-fibreglass hybrid casts should be considered for orthopedic use on the basis of their strength, stiffness, weight and cost, combined with their acknowledged advantages of good moulding ability and water resistance.

Mesh:

Substances:

Year:  2000        PMID: 11045095      PMCID: PMC3695143     

Source DB:  PubMed          Journal:  Can J Surg        ISSN: 0008-428X            Impact factor:   2.089


  6 in total

Review 1.  Serial elongation derotation flexion casting in children with infantile and juvenile scoliosis.

Authors:  Federico Canavese; Alain Dimeglio
Journal:  Ann Transl Med       Date:  2020-01

2.  The Impact of Different Application Techniques on Fiberglass Casts: A Mechanical Experimental Study.

Authors:  Mehmet Yalçınozan; Enes Sarı
Journal:  Indian J Orthop       Date:  2020-07-02       Impact factor: 1.251

3.  Lower limb intracast pressures generated by different types of immobilisation casts.

Authors:  Salma Chaudhury; Alexandra Hazlerigg; Anuhya Vusirikala; Joseph Nguyen; Stuart Matthews
Journal:  World J Orthop       Date:  2017-02-18

4.  Physical and Clinical Evaluation of Hip Spica Cast applied with Three-slab Technique using Fibreglass Material.

Authors:  K M Bitar; M E Ferdhany; E I Ashraf; A Saw
Journal:  Malays Orthop J       Date:  2016-11

5.  Comparison of two different ways to apply a circular plaster cast for distal radius fractures: biomechanical study.

Authors:  Alejandro Espejo-Reina; María T Carrascal-Morillo; Alberto D Delgado-Martínez
Journal:  J Orthop Surg Res       Date:  2021-01-30       Impact factor: 2.359

6.  Test of Strength: Figure-of-Eight versus Spiral Wrapping Technique for Fiberglass Casts.

Authors:  Blake K Montgomery; Hunter W Storaci; Nicole A Segovia; Jeffrey Young
Journal:  Cureus       Date:  2020-04-26