| Literature DB >> 11043789 |
Abstract
Gas chromatography coupled to positive and negative ion chemical ionisation mass spectrometry was evaluated for the determination of free volatile fatty acids (VFAs) from aqueous samples by headspace solid-phase microextraction. Negative ion chemical ionisation in the selected ion monitoring mode using ammonia as reagent gas provided acceptable sensitivity and the highest selectivity for the determination of C2-C7 fatty acids using a polydimethylsiloxane-Carboxen fibre. Detection limits in the range of 150 microg l(-1) for acetic acid and from 2 to 6 microg l(-1) for the remaining carboxylic acids were achieved. The reproducibility of the method was between 9 and 16%. The developed analytical procedure was applied to the analysis of VFAs in raw sewage. The absence of interfering peaks provided a more accurate determination of acetic, propionic, butyric and isovaleric acids than a similar analytical scheme but using a flame ionisation detector.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11043789 DOI: 10.1016/s0021-9673(00)00655-5
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759