Literature DB >> 11035865

Validation of A&D UA-767 device for the self-measurement of blood pressure.

A N Rogoza1, T S Pavlova, M V Sergeeva.   

Abstract

BACKGROUND: The validation of self-measurement devices has been recommended. Automatic monitor A&D UA-767 (A&D Company, Ltd, Tokyo, Japan) is well known and widely used, but not tested according to the Association for Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) recommendations.
OBJECTIVE: To perform a clinical validation for use by adults of the A&D UA-767 device according to the criteria of the AAMI and a modified BHS protocol.
METHODS: The test concerned 101 subjects (57 men and 44 women) aged 15-85 years with arm circumferences of 22-39 cm, a systolic blood pressure (SBP) range of 89-206 mmHg, and a diastolic blood pressure (DBP) range of 53-122 mmHg. For each subject, three readings of the UA-767 were compared with simultaneous auscultatory measurements by two trained independent observers who used a mercury manometer and dual stethoscope. The results were graded according to the BHS (1990 and 1993) and AAMI recommendations.
RESULTS: Observers showed close agreement, with mean differences of 1.1+/-2.4 mmHg for SBP and -0.7+/-2.0 mmHg for DBP. The proportion of values agreeing to within 5, 10 and 15 mmHg were 93, 100, and 100% for SBP and 97, 100, and 100% for DBP for the two observers. The sphygmomanometer measurements were 132+/-24/79+/-14 mmHg (mean+/-SD). The average difference between the mercury sphygmomanometer and A&D UA-767 readings for SBP and DBP were, respectively, -0.4+/-5.4 and -0.4+/-4. 8 mmHg. The proportion of values agreeing to within 5,10, and 15 mmHg were 82, 94, and 98% for SBP and 80, 95, and 98% for DBP for the observers and device (A/A grade for BHS).
CONCLUSIONS: For an adult population, the A&D UA-767 device for the self-measurement of blood pressure satisfied the AAMI criteria, achieved a BHS grade of A/A and can therefore be recommended for monitoring blood pressure in home and clinical conditions by patients with mild-to-moderate arterial hypertension.

Entities:  

Mesh:

Year:  2000        PMID: 11035865     DOI: 10.1097/00126097-200008000-00006

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  28 in total

1.  Development and evaluation of a hypertension knowledge test for Korean hypertensive patients.

Authors:  Hae-Ra Han; Kitty Chan; Heejung Song; Tam Nguyen; Jong-Eun Lee; Miyong T Kim
Journal:  J Clin Hypertens (Greenwich)       Date:  2011-07-14       Impact factor: 3.738

2.  Waiting a few extra minutes before measuring blood pressure has potentially important clinical and research ramifications.

Authors:  S B Nikolic; W P Abhayaratna; R Leano; M Stowasser; J E Sharman
Journal:  J Hum Hypertens       Date:  2013-05-30       Impact factor: 3.012

3.  Pragmatic Method Using Blood Pressure Diaries to Assess Blood Pressure Control.

Authors:  James E Sharman; Leigh Blizzard; Wojciech Kosmala; Mark R Nelson
Journal:  Ann Fam Med       Date:  2016 Jan-Feb       Impact factor: 5.166

4.  The population-based prevalence of hypertension and correlates of blood pressure among Australian children.

Authors:  Nicholas G Larkins; Armando Teixeira-Pinto; Siah Kim; David P Burgner; Jonathan C Craig
Journal:  Pediatr Nephrol       Date:  2019-02-18       Impact factor: 3.714

5.  The effect of a community-based self-help multimodal behavioral intervention in Korean American seniors with high blood pressure.

Authors:  Kim B Kim; Hae-Ra Han; Boyun Huh; Tam Nguyen; Hochang Lee; Miyong T Kim
Journal:  Am J Hypertens       Date:  2014-03-26       Impact factor: 2.689

6.  Electrocardiographic reference values for a population of older adults in sub-Saharan Africa.

Authors:  Matthew J Dewhurst; Luigi Y Di Marco; Felicity Dewhurst; Philip C Adams; Alan Murray; Golda P Orega; Julius C Mwita; Richard W Walker; Philip Langley
Journal:  Ann Noninvasive Electrocardiol       Date:  2013-09-09       Impact factor: 1.468

7.  A comparison of blood pressure measurement over a sleeved arm versus a bare arm.

Authors:  Grace Ma; Norman Sabin; Martin Dawes
Journal:  CMAJ       Date:  2008-02-26       Impact factor: 8.262

8.  Knowledge, beliefs, and behaviors about hypertension control among middle-aged Korean Americans with hypertension.

Authors:  Hae-Ra Han; Kim B Kim; Jeonghee Kang; Seonghee Jeong; Eun-Young Kim; Miyong T Kim
Journal:  J Community Health       Date:  2007-10

9.  Perceptions of coronary heart disease risk in Korean immigrants with type 2 diabetes.

Authors:  Sarah Choi; Sally Rankin; Anita Stewart; Roberta Oka
Journal:  Diabetes Educ       Date:  2008 May-Jun       Impact factor: 2.140

10.  Prospective change in renal volume and function in children with ADPKD.

Authors:  Melissa A Cadnapaphornchai; Kim McFann; John D Strain; Amirali Masoumi; Robert W Schrier
Journal:  Clin J Am Soc Nephrol       Date:  2009-04       Impact factor: 8.237

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.