Clinical evaluation of the first medical whole blood, point-of-care testing device for detection of myocardial infarction.

F S Apple1, F P Anderson, P Collinson, R L Jesse, M C Kontos, M A Levitt, E A Miller, M M Murakami.   

Abstract

BACKGROUND: Validation of whole blood, point-of-care testing devices for monitoring cardiac markers to aid clinicians in ruling in and ruling out myocardial infarction (MI) is necessary for both laboratory and clinical acceptance.
METHODS: This study evaluated the clinical diagnostic sensitivity and specificity of the First Medical Cardiac Test device operated by nursing and laboratory personnel that simultaneously measures cardiac troponin I (cTnI), creatine kinase (CK) MB, myoglobin, and total CK on the Alpha Dx analyzer in whole blood for detection of MI. Over a 6-month period, 369 patients initially presenting to the emergency department with chest pain were evaluated for MI using modified WHO criteria. Eighty-nine patients (24%) were diagnosed with MI.
RESULTS: In whole blood samples collected at admission and at 3- to 6-h intervals over 24 h, ROC curve-determined MI decision limits were as follows: cTnI, 0.4 microgram/L; CKMB, 7.0 microgram/L; myoglobin, 180 microgram/L; total CK, 190 microgram/L. Based on peak concentrations within 24 h after presentation, the following sensitivities (+/- 95% confidence intervals) were found: cTnI, 93% +/- 5.5%; myoglobin, 81% +/- 9.7%; CKMB, 90% +/- 6.3%; total CK, 86% +/- 7.5%. Sensitivities were maximal at >90% for both cTnI and CKMB at >12 h in MI patients, without differences between ST-segment elevation and non-ST-segment elevation MI patients.
CONCLUSIONS: The First Medical point-of-care device provides cardiac marker assays that can be used by laboratories and clinicians in a variety of hospital settings for ruling in and ruling out MI.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11017938

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

1.  Myoglobin immunoassay based on metal particle-enhanced fluorescence.

Authors:  Evgenia G Matveeva; Zygmunt Gryczynski; Joseph R Lakowicz
Journal:  J Immunol Methods       Date:  2005-07       Impact factor: 2.303

2.  Management strategies for patients with low-risk chest pain in the emergency department.

Authors:  Maame Yaa A B Yiadom; Joshua M Kosowsky
Journal:  Curr Treat Options Cardiovasc Med       Date:  2011-02

3.  Use of portable blood physiology point-of-care devices for basic and applied research on vertebrates: a review.

Authors:  Lauren J Stoot; Nicholas A Cairns; Felicia Cull; Jessica J Taylor; Jennifer D Jeffrey; Félix Morin; John W Mandelman; Timothy D Clark; Steven J Cooke
Journal:  Conserv Physiol       Date:  2014-04-04       Impact factor: 3.079

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.