Effect of shaving the chest of hirsute subjects on transthoracic impedance to self-adhesive defibrillation electrode pads.

J W Bissing1, R E Kerber.   

Abstract

We determined transthoracic impedance (TTI) before and after shaving in 9 hirsute and 11 nonhirsute subjects. TTI in hirsute subjects was mean +/- SE 162 +/- 11 versus 103 +/- 6 ohms in nonhirsute subjects (p <0.01); TTI decreased to 105 +/- 3 ohms after shaving the hirsute chests (p <0.01).

Mesh:

Year:  2000        PMID: 11009290     DOI: 10.1016/s0002-9149(00)01026-2

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  2 in total

Review 1.  How good is your defibrillation technique?

Authors:  Daniel M Sado; Charles D Deakin
Journal:  J R Soc Med       Date:  2005-01       Impact factor: 18.000

Review 2.  A Systematic Review of the Transthoracic Impedance during Cardiac Defibrillation.

Authors:  Yasmine Heyer; Daniela Baumgartner; Christian Baumgartner
Journal:  Sensors (Basel)       Date:  2022-04-06       Impact factor: 3.576

  2 in total