Influence of sulphuric acid contaminants on Fricke dosimetry.

A Palm1, O Mattsson.   

Abstract

The sulphuric acid used for the preparation of the Fricke dosimeter solution may contain trace impurities that can affect the yield of ferric ions. Two methods, pre-irradiation or oxidation with hydrogen peroxide, have been proposed to reduce the influence of these impurities. Fricke users sometimes omit this treatment. In the present work Fricke solutions prepared from six different brands and qualities of sulphuric acid were compared in order to study any influences of the acid on the ferric ion yield. It was shown that the use of analytical grade sulphuric acid from one manufacturer resulted in a reduction of the ferric ion yield of about 5% at an absorbed dose of approximately 20 Gy. If this solution were to be used for an absolute dose determination together with epsilon(m) G values from the literature the absorbed dose would be underestimated by the same amount.