Literature DB >> 11001508

Detection of fatigue in the isometric electrical activation of paralyzed hand muscles of persons with tetraplegia.

J M Heasman1, T R Scott, V A Vare, R Y Flynn, C R Gschwind, J W Middleton, S B Rutkowski.   

Abstract

Paralyzed muscle fatigue is the eventual depression of force due to either prolonged or repetitive electrical stimulation of motor units. The robustness and safety of future functional electrical stimulation (FES) systems will rely on their ability to detect the onset of muscle fatigue. The relative degree of muscle activation can be estimated by monitoring the M-wave. The aim of this study was to test a proposed method of quantitative fatigue assessment that detects muscle force output and its corresponding M-wave measured concurrently. The detection of force and M-wave concurrently allows any reduction in muscle force output to be attributed to either changes in the fatigue state of the stimulated muscle or changes in the degree of stimulus activation of that muscle. The fatigue assessment scheme can thereby accommodate the corresponding changes in muscle force caused by an alteration in the stimulation intensity during fatigue. The Extensor Digitorum Communis (EDC), Extensor Pollicis Longus (EPL), and Flexor Pollicis Longus (FPL) muscles of two C5/C6 tetraplegic men were studied. Stimulation recruitment tests over the pulsewidth range from 0 to 200 micros, were performed at intervals during 20 min of maximal stimulation (200 micro/s). Muscle force correlated to the M-wave parameter, second phase area, with mean correlation coefficients of greater than 0.82, when the muscle was in either a nonfatigued or fatiguing state. After the initial force, likely to be primarily due to the fast glycolytic (FG) motor units, had declined the M-wave demonstrated only minor changes throughout the fatigue of muscle force during 20 min of constant maximal stimulation. The second phase area and root-mean-square (rms) of the M-wave [see Fig. 2(a) reflected muscle activation during modulated stimulation and also remained relatively constant during the fatigue-related force decline when the muscle was stimulated at a constant intensity. This detection of M-wave parameters satisfies the defined requirement for a myoelectric parameter that indicates electrical activation, but is relatively invariant to muscular fatigue. Index Terms-Electrical stimulation, electromyography (EMG), functional electrical stimulation (FES), muscle fatigue, spinal cord injury, tetraplegia.

Entities:  

Mesh:

Year:  2000        PMID: 11001508     DOI: 10.1109/86.867870

Source DB:  PubMed          Journal:  IEEE Trans Rehabil Eng        ISSN: 1063-6528


  3 in total

Review 1.  The effectiveness of FES-evoked EMG potentials to assess muscle force and fatigue in individuals with spinal cord injury.

Authors:  Morufu Olusola Ibitoye; Eduardo H Estigoni; Nur Azah Hamzaid; Ahmad Khairi Abdul Wahab; Glen M Davis
Journal:  Sensors (Basel)       Date:  2014-07-14       Impact factor: 3.576

2.  Evoked EMG versus muscle torque during fatiguing functional electrical stimulation-evoked muscle contractions and short-term recovery in individuals with spinal cord injury.

Authors:  Eduardo H Estigoni; Che Fornusek; Nur Azah Hamzaid; Nazirah Hasnan; Richard M Smith; Glen M Davis
Journal:  Sensors (Basel)       Date:  2014-12-03       Impact factor: 3.576

3.  Evoked Electromyographically Controlled Electrical Stimulation.

Authors:  Mitsuhiro Hayashibe
Journal:  Front Neurosci       Date:  2016-07-14       Impact factor: 4.677

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.