Literature DB >> 10999697

Bronchial perforation after closed-tube endotracheal suction.

A Thakur1, T Buchmiller, J Atkinson.   

Abstract

Iatrogenic bronchial perforation is an uncommon complication in intubated premature infants. The authors present 2 cases of bronchial perforation caused by closed-tube endotracheal suction catheters. These presented in the first 3 days of life with pneumothorax requiring chest tube placement. Medical management was unsuccessful because of the high-volume air leak present and surgical intervention was required. A bronchial perforation of the right lower bronchus was identified in both cases and was repaired in the second case. This represents the first report in the literature of iatrogenic bronchial perforation in premature infants by closed-tube endotracheal suctioning catheters. Bronchial perforation requires early recognition and prompt repair if successful outcome is to be obtained.

Entities:  

Mesh:

Year:  2000        PMID: 10999697     DOI: 10.1053/jpsu.2000.9332

Source DB:  PubMed          Journal:  J Pediatr Surg        ISSN: 0022-3468            Impact factor:   2.545


  4 in total

1.  Unusual complication of reusable suction catheter during rigid bronchoscopy.

Authors:  Churunal K Hari; Tim Petheram; Richard Garth
Journal:  Eur Arch Otorhinolaryngol       Date:  2007-07-04       Impact factor: 2.503

2.  Pulmonary air leak syndrome associated with traumatic nasopharyngeal suctioning.

Authors:  Maria Chitty Lopez; Elza Pollak-Christian; Michael Keenaghan
Journal:  BMJ Case Rep       Date:  2016-10-28

3.  Successful Surgical Management of Airway Perforation in Preterm Infants.

Authors:  Jason Gien; Richard J Ing; Mark D Twite; David Campbell; Max Mitchell; John P Kinsella
Journal:  J Pediatr Surg Case Rep       Date:  2014-02-01

4.  Conservative Treatment of Lung Perforation Secondary to Retained Catheter in an Extremely Low-Birth-Weight Premature Infant.

Authors:  Mehmet Surhan Arda; Miller C Hamrick; Timothy D Kane
Journal:  European J Pediatr Surg Rep       Date:  2015-06-26