Contamination of piped medical gas supply with water.

H Hay1.   

Abstract

The failure of anaesthetic equipment as a result of maintenance is extremely rare. The ingress of water into the flowmeters of an anaesthetic machine from the piped medical air supply is reported and is possibly unique. The piped medical air supply was open to the atmosphere during maintenance. Water condensed in the gas pipeline and this was not noticed during subsequent testing. Water was seen leaking from the orthopaedic air tools used for surgery but was assumed to be from the autoclaving process. Later the same day, when medical air from the piped source was used as part of the gas mixture for a general anaesthetic, water was seen filling the barrel of the flowmeter air control valve. This could have had far-reaching and dangerous consequences for the patient, which were fortunately averted.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 10998035     DOI: 10.1046/j.1365-2346.2000.00702.x

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  4 in total

1.  Water in a nitrous oxide flowmeter.

Authors:  Virendra K Arya; Vikramjeet Arora
Journal:  Indian J Anaesth       Date:  2010-03

2.  Sudden Appearance of Water in Flowmeter During Air/Oxygen and Sevoflurane Anaesthesia.

Authors:  Tünay Kandemir; Selda Muslu; Erbin Kandemir
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-09-09

3.  Critical incident is a possibility with water in flowmeter.

Authors:  M Manjuladevi; Ks Vasudeva Upadhyaya; Ps Sathyanarayana; Rashmi Rani; Viraj Shah
Journal:  Indian J Anaesth       Date:  2014 Nov-Dec

4.  Water contamination of the central air supply: Out of sight, out of mind!

Authors:  Geeta Kamal; Anju Gupta; Sapna Bathla; Aikta Gupta
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2018 Jan-Mar
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.