C Berndt1, U Müller, F Bergmann, U Schmitt, R Kaiser, C Müller.
Abstract
HIV RNA was quantified in blood plasma from 209 patients and in control specimen comparing the NucliSens HIV-1 QT test (Organon Teknika), which is based on the nucleic acid sequence amplification procedure, and the Quantiplex 3.0 test (Bayer), which uses hybridization signal enhancement by branched DNA (bDNA) probes. A highly significant correlation (P=0.01) was found between the two methods with 88% of the samples showing similar results. In cases of discrepant findings, higher virus load was observed with either test (14xNASBA>bDNA; 12xbNDA>NASBA). Differences could neither be related to clinical features nor to divergent virus subtypes. Standard preparations containing 35000 and 222000 copies were quantified with intra-assay coefficients of variation of <20% using both methods. A preparation of 192 copies was measured with lower precision by both tests, yet was detected more reliably by the bDNA method.Entities: