OBJECTIVES: To compare the diagnostic efficacy of Tuned Aperture Computed Tomography (TACT) with conventional imaging modalities for detection of fractures of the mandible.
METHODS: Fractures were induced using blunt trauma in human defleshed mandibles. Conventional extra-oral and indirect digital images, unprocessed TACT images reconstructed from eight and 16 basis images (BI) and iteratively restored TACT images from eight and 16 BI were used for fracture evaluation. Twelve observers recorded their diagnoses using a five-point confidence rating scale. The data were analysed using ROC curve analysis.
RESULTS: Significant differences were found (P < 0.0001) in the areas under the curve (Az): film, 0.6954; digital images 0.6169; TACT unprocessed images using 8 BI, 0.7420; TACT unprocessed images using 8 BI, 0.7667; TACT unprocessed images using 16 BI, 0.7730; TACT iteratively restored images using 16 BI, 0.8143. No observer-based differences were found. Fractures in the condylar and coronoid regions were more difficult to detect than those in the ramus and body of the mandible.
CONCLUSIONS: Iteratively restored TACT images generated using 16 BI had a superior diagnostic performance to all the other imaging modalities. Studies are in order to evaluate its in vivo potential.

Entities:  

Mesh:

Year:  2000        PMID: 10980566     DOI: 10.1038/sj/dmfr/4600548

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  3 in total

1.  Digital tomosynthesis-experiences with a new imaging device for the dental field.

Authors:  Christoph M Ziegler; Manfred Franetzki; Tina Denig; Joachim Mühling; Stefan Hassfeld
Journal:  Clin Oral Investig       Date:  2003-02-18       Impact factor: 3.573

2.  Stationary intraoral digital tomosynthesis using a carbon nanotube X-ray source array.

Authors:  J Shan; A W Tucker; L R Gaalaas; G Wu; E Platin; A Mol; J Lu; O Zhou
Journal:  Dentomaxillofac Radiol       Date:  2015-06-19       Impact factor: 2.419

3.  Characterization and preliminary imaging evaluation of a clinical prototype stationary intraoral tomosynthesis system.

Authors:  Christina R Inscoe; Enrique Platin; Sally M Mauriello; Angela Broome; Andre Mol; Laurence R Gaalaas; Michael W Regan Anderson; Connor Puett; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2018-10-17       Impact factor: 4.071

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.