| Literature DB >> 10977780 |
S Schmaldienst1, A Goldammer, S Spitzauer, K Derfler, W H Hörl, P Knöbl.
Abstract
A regimen of local anticoagulation of an immunoadsorption device was studied. The extracorporeal circuit was anticoagulated with citrate (5.5%) and a continuous infusion of heparin (2,000 U/h or 1,500 U/h), which was neutralized by a continuous infusion of protamine chloride (75% of the heparin dose) before reinfusion in 23 patients treated with low-density lipoprotein or immunoglobulin apheresis. Sufficient anticoagulation of the extracorporeal circuit was obtained (activated partial thromboplastin time [APTT] > 180 seconds; thrombin time [TT] > 120 seconds; anti-Xa activity, 1.05 +/- 0.21 U/mL) during the entire treatment of 190 minutes, whereas coagulation parameters in the patients' blood stayed within the normal range. In a control group without heparin neutralization, full systemic anticoagulation of the patients occurred (APTT, 157.8 +/- 30.6 seconds; TT, 119.8 +/- 0.4 seconds; anti-Xa activity, 0.88 +/- 0.21 U/mL). No side effects or clotting of the system were observed. Our data show that this regimen of local anticoagulation is a safe protocol for extracorporeal circulation without exposing the patients to anticoagulants.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10977780 DOI: 10.1053/ajkd.2000.9789
Source DB: PubMed Journal: Am J Kidney Dis ISSN: 0272-6386 Impact factor: 8.860