K Racaityte1, K K Unger, D Lubda, K S Boos.
Abstract
A novel restricted access cation exchanger with sulphonic acid groups at the internal surface was proven to be highly suitable in the sample clean up of peptides on-line coupled to HPLC-electrospray ionization (ESI)-MS. Neuropeptide Y (NPY) and several of its fragments in plasma were subjected to the sample clean-up procedure. The peptides were eluted by a step gradient from the restricted access column, applying 10 mM phosphate buffer pH 3.5 from 5 to 20% (v/v) of acetonitrile with 1 M NaCl and transferred to a Micra ODS II column (33x4.6 mm). The separation of the peptides and their fragments was performed by a linear gradient from 20 to 60% (v/v) acetonitrile in water with 0.1% formic acid and 0.01% trifluoroacetic acid in 4 min at a flow-rate of 0.75 ml/min. An integrated and completely automated system composed of sample clean up-HPLC-ESI-MS was used to analyze real life samples. The sample volumes ranged between 20 and 100 microl. Peaks due to the fragments NPY 1-36, 3-36 and 13-36 in porcine plasma were identified by ESI-MS. The limit of detection was in the 5 nmol/ml range. The total analysis required 21 min and allowed the direct injection of plasma.Entities:
Mesh:
Substances:
Year: 2000 PMID: 10976801 DOI: 10.1016/s0021-9673(00)00639-7
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759