A new interference nail fixation for anterior cruciate ligament soft tissue grafts.

A U Turhan1, O Aynaci, H Aydin, A Biyiklioğlu.   

Abstract

We present a new device for femoral and tibial fixation in anterior cruciate ligament reconstruction using soft tissue. We evaluated the method biomechanically in pullout tests using human cadaver knee joints. Fixation strengths ranged from 300 +/- 60 to 600 +/- 50 N (mean 400 +/- 50 N). This corresponds to loads in the graft during aggressive rehabilitation although cadaveric specimens were used. With respect to primary fixation strength this implant is a reasonable alternative for anterior cruciate ligament graft fixation.

Entities:  

Mesh:

Year:  2000        PMID: 10975261     DOI: 10.1007/s001670000119

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  1 in total

1.  Initial fixation strength of interference nail fixation for anterior cruciate ligament reconstruction with patellar tendon graft (experimental study).

Authors:  H Aydin; S Kerimoğlu; A Citlak; A U Turhan
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-07-31       Impact factor: 4.342